7. CALIBRATION

This function adjusts the balance for accurate weighing. Perform a calibration in the following cases.

‰ When the balance is first used.

Calibration (CAL) switch

‰ When the balance has been moved. ‰ For regular calibration.

Calibration (CAL) switch cover

7-1. Calibration using a weight

‰Prepare the calibration weight (sold separately) in advance.

1. Warm up the balance for at least half an hour with

Press and hold

the CAL switch.

nothing on the pan.

 

 

 

 

 

2. Press and hold the calibration (CAL) switch until

Cal

Cal

appears, and release the switch.

Release the CAL switch.

 

 

 

 

3. The balance displays

Cal 0 .

 

To change the calibration weight value, proceed to

Cal 0

step 4.

 

 

 

To use the calibration weight value in the balance

 

memory, proceed to step 5.

 

4. Press the

SAMPLE

key. The display shows the

060000.

calibration weight value in “gram” that is stored in the

balance. Use the following keys to change the value.

Set the weight using

 

 

 

 

SAMPLE

key

To

select the digit blinking to

the relevant keys.

 

change.

RE-ZERO

key

To set the value of the digit

 

 

 

selected.

 

key

To store the value and return to

PRINT

 

 

 

step 3.

Cal 0

UNITS

key

To cancel the value and return to

 

step 3.

14

Image 16
Contents 1WMPD4001917 This manual and Marks Compliance with FCC rulesContents Maintenance Introduction UnpackingNET indicator Part Names and FunctionsSetting up your balance Setting UPPower source Breeze break Installing batteriesStorage Operation Turn the power on and OFFLCD backlight Units Conversion tableSelecting a weighing unit NETBasic operation Counting mode pcs Selecting the counting modeCounting mode using the Acai function Storing the sample unitPercent mode % 100100 0 % 1000. %Setting example Selecting a comparison modeComparator FuncCp Hi 000000 001234Cp lo 000000 001230Calibration using a weight CalibrationCal 060000Cal 50000 GlpCal 97985 Gravity acceleration correctionEntering the function setting mode FunctionsCp Hi Key operationACai Func PoffEnd Cp Hi End id UnitStoring weighing units Function list Func ACai UminSmpl LdinEJ-03 Installation OptionsEJ-02 USB Interface EJ-03 RS-232C serial interfaceInterface specifications EIA RS-232CData output mode Data formatCommand mode Command listEJ-07 / EJ-08 Installation EJ-07 / EJ-08 UnderhookEJ-11 Breeze break EJ-12 Carrying caseEJ-13 Density Determination Kit EJ-13 InstallationDensity specific gravity measurement Change the function table 000Example of density measurement 2000 1751 8801Setting the ID number ID Number and GLPEnd id func Display character setData format for calibration report Output exampleData format for calibration test report Calibration test using a weight TitleEnd Block Title block and End block20000 20000 EndMaintenance Error codesOverload error Range over notice Unit weight error Sample quantity noticeError1 Cal e -Cal eLb0 Hb0 Hb1 Lb1Specifications EJ seriesOther weighing units Dimensions OptionsGravity Acceleration MAP Values of gravity at various locationsWorld map