A&D EJ-440, EJ-6100 manual Counting mode pcs, Selecting the counting mode, Storing the sample unit

Page 12
25
ACAI indicator

5-6. Counting mode (pcs)

Weighs the sample pieces and calculate its unit weight. Using the sample unit weight, the scale counts the number of parts or items.

‰As for the minimum unit weight acceptable, see the function setting “Umin”.

Selecting the counting mode

1. Press the

UNITS

key to select

.

( :pieces)

Storing the sample unit

2.Press the SAMPLE key to enter the sample unit weight storing mode.

3.To select the number of samples, press the

SAMPLE key. It may be set to 5, 10, 25, 50, or 100.

0

10 -

Each pressing switches the number of samples.

 

25 -

4. Place a tare container on the weighing pan, and

Container

press the RE-ZERO key. Confirm that the

 

right side of the number of samples shows zero.

 

 

Weighing pan

Confirm the display

 

5.Place the number of samples specified on the pan. In this example, 25 pieces.

6.Press the PRINT key to calculate and store the unit weight. Remove the sample. The balance is set to count objects with this unit weight.

ZERO

25 0

 

Samples (25 pcs)

To store

Counting the objects

25

7. Place the objects to be counted on the pan.

Counting mode using the ACAI function

ACAITM (Automatic Counting Accuracy Improvement) is a function that improves the accuracy of the unit weight by increasing the number of samples as the counting process proceeds.

8.If a few more samples are added, the ACAI indicator is displayed. (To prevent an error, add three or more. The ACAI indicator will not be displayed if overloaded.)

ACAI

10

Image 12
Contents 1WMPD4001917 This manual and Marks Compliance with FCC rulesContents Maintenance Introduction UnpackingNET indicator Part Names and FunctionsSetting UP Setting up your balancePower source Installing batteries Breeze breakStorage Turn the power on and OFF OperationLCD backlight Units Conversion tableNET Selecting a weighing unitBasic operation Counting mode pcs Selecting the counting modeCounting mode using the Acai function Storing the sample unitPercent mode % 100100 0 % 1000. %Setting example Selecting a comparison modeComparator FuncCp Hi 000000 001234Cp lo 000000 001230Calibration using a weight CalibrationCal 060000Cal 50000 GlpCal 97985 Gravity acceleration correctionEntering the function setting mode FunctionsCp Hi Key operationFunc Poff ACaiEnd Cp Hi Unit End idStoring weighing units Function list Func ACai UminSmpl LdinEJ-03 Installation OptionsEJ-02 USB Interface EJ-03 RS-232C serial interfaceInterface specifications EIA RS-232CData output mode Data formatCommand mode Command listEJ-07 / EJ-08 Installation EJ-07 / EJ-08 UnderhookEJ-11 Breeze break EJ-12 Carrying caseEJ-13 Installation EJ-13 Density Determination KitDensity specific gravity measurement 000 Change the function tableExample of density measurement 2000 1751 8801Setting the ID number ID Number and GLPEnd id func Display character setOutput example Data format for calibration reportData format for calibration test report Calibration test using a weight TitleEnd Block Title block and End block20000 20000 EndMaintenance Error codesOverload error Range over notice Unit weight error Sample quantity noticeError1 Cal e -Cal eLb0 Hb0 Hb1 Lb1Specifications EJ seriesOther weighing units Dimensions OptionsGravity Acceleration MAP Values of gravity at various locationsWorld map