5.At step 3, pressing the PRINT key to weigh the
With nothing on the pan
To store
6.Place the calibration weight with the same value as displayed on the pan. Press the PRINT key to weigh it. Do not touch the pan during weighing.
7. | end | appears. |
Remove the weight from the pan.
When the GLP output (function setting “info 1” or
“info 2”) is selected, | glp |
| is displayed. The | |||
calibration report is output and |
| end | appears | |||
again. |
|
|
|
| ||
Press the CAL switch or | UNITS |
| key to return to | |||
the weighing mode. |
|
|
|
|
Cal 0
50000.
Calibration weight
To store
50000.
0end000
glp
GLP output
0end000
Returns to
the weighing mode
Note
The value set in step 4 is stored in memory even after the power is switched off.
If the balance is to be moved to other places, set the gravity acceleration value of the area where the calibration using a weight is to be done, and calibrate the balance according to the procedure above. See the next section to set the value.
15