A&D EJ-6100, EJ-440, EJ-3000, EJ-200, EJ-4100, EJ-120 100 0 %, 1000. %, 758. %, Percent mode %

Page 13

9.The balance re-calculates the unit weight while the ACAI indicator is blinking. Do not touch the balance or samples on the pan until the ACAI indicator turns off.

10.Counting accuracy is improved when the ACAI indicator turns off. Each time the above operation is performed, a more accurate unit weight will be obtained. There is no definite upper limit of ACAI range for the number of samples exceeding 100. Try to add the similar number of samples as displayed.

5-7. Percent mode (%)

Displays the weighing value in percentage compared to the reference (100%) weight.

Selecting the percent mode

1. Press the UNITS key to select %. (%:percent)

Storing the reference (100%) weight

2.Press the SAMPLE key to enter the reference weight storing mode.

00.

100 -

%

%

3. Press the

RE-ZERO

key to display 100 0%.

4. Place the sample

to be set as the reference

ZERO

 

100 0 %

 

 

Weighing pan

 

 

weight on the pan.

 

 

 

 

 

 

 

5.Press the PRINT key to store the reference weight. Remove the sample.

Reading the percentage

6.Place the object to be compared to the reference weight on the pan. The displayed percentage is based on 100% of the reference weight.

Sample corresponding to the 100% weight

To store

1000. %

Object to be compared

758. %

11

Image 13
Contents 1WMPD4001917 Compliance with FCC rules This manual and MarksContents Maintenance Unpacking IntroductionPart Names and Functions NET indicatorSetting up your balance Setting UPPower source Breeze break Installing batteriesStorage Operation Turn the power on and OFFLCD backlight Conversion table UnitsSelecting a weighing unit NETBasic operation Selecting the counting mode Counting mode pcsCounting mode using the Acai function Storing the sample unit100 Percent mode %100 0 % 1000. %Selecting a comparison mode Setting exampleComparator Func001234 Cp Hi 000000Cp lo 000000 001230Calibration Calibration using a weightCal 060000Glp Cal 50000Gravity acceleration correction Cal 97985Functions Entering the function setting modeCp Hi Key operationACai Func PoffEnd Cp Hi End id UnitStoring weighing units Function list Umin Func ACaiSmpl LdinOptions EJ-03 InstallationEJ-02 USB Interface EJ-03 RS-232C serial interfaceEIA RS-232C Interface specificationsData format Data output modeCommand list Command modeEJ-07 / EJ-08 Underhook EJ-07 / EJ-08 InstallationEJ-11 Breeze break EJ-12 Carrying caseEJ-13 Density Determination Kit EJ-13 InstallationDensity specific gravity measurement Change the function table 000Example of density measurement 1751 8801 2000ID Number and GLP Setting the ID numberEnd id func Display character setData format for calibration report Output exampleData format for calibration test report Title Calibration test using a weightEnd Block Title block and End block20000 End 20000Error codes MaintenanceOverload error Range over notice Unit weight error Sample quantity noticeCal e -Cal e Error1Lb0 Hb0 Hb1 Lb1EJ series SpecificationsOther weighing units Options DimensionsValues of gravity at various locations Gravity Acceleration MAPWorld map