9.The balance
10.Counting accuracy is improved when the ACAI indicator turns off. Each time the above operation is performed, a more accurate unit weight will be obtained. There is no definite upper limit of ACAI range for the number of samples exceeding 100. Try to add the similar number of samples as displayed.
5-7. Percent mode (%)
Displays the weighing value in percentage compared to the reference (100%) weight.
Selecting the percent mode
1. Press the UNITS key to select %. (%:percent)
Storing the reference (100%) weight
2.Press the SAMPLE key to enter the reference weight storing mode.
00.
100 -
%
%
3. Press the
RE-ZERO
key to display 100 0%.
4. Place the sample | to be set as the reference | ZERO |
| 100 0 % |
| ||||
| Weighing pan | |||
|
| |||
weight on the pan. |
|
|
| |
|
|
|
|
5.Press the PRINT key to store the reference weight. Remove the sample.
Reading the percentage
6.Place the object to be compared to the reference weight on the pan. The displayed percentage is based on 100% of the reference weight.
Sample corresponding to the 100% weight
To store
1000. %
Object to be compared
758. %
11