A&D Essential Maintenance and Error Code Guide for EJ-610 Scale Model

Page 36

11. MAINTENANCE

11-1. Notes on maintenance

‰Do not disassemble the balance. Contact your local A&D dealer if your balance needs service or repair.

‰Please use the original package for transportation.

‰Do not use organic solvents to clean the balance. Use a warm lint free cloth dampened with a mild detergent.

‰To clean around the pan support, refer to the drawings below.

Weighing pan

Weighing pan

Raised portion

Wind shield

Raised

 

 

 

portion

 

Fixing frame

 

Pan support

Pan support

EJ-120

/ EJ-200 / EJ-300

EJ-1500

/ EJ-2000 / EJ-3000

EJ-410

/ EJ-610

EJ-4100

/ EJ-6100

11-2. Error codes

Overload error

eeeeee

Range over notice

--e

Unit weight error

lo

Sample quantity notice

10 -

25 -

Warning to indicate that an object beyond the balance capacity has been placed on the pan. Remove the object from the pan.

This will be shown if the weight sensor receives strong force upward. Check if there is anything sandwiched around the weighing pan. There is a possibility that the weight sensor itself may have a failure.

The sample weight is too light to set the unit weight in the counting mode or 100% reference weight in the % mode.

When sample weight is light and the counting error could become large, the balance will request you to use larger number of samples. Place the displayed number of samples on the pan and press the PRINT key to store

34

Image 36
Contents 1WMPD4001917 This manual and Marks Compliance with FCC rulesContents Maintenance Introduction UnpackingNET indicator Part Names and FunctionsSetting UP Setting up your balancePower source Installing batteries Breeze breakStorage Turn the power on and OFF OperationLCD backlight Units Conversion tableNET Selecting a weighing unitBasic operation Counting mode pcs Selecting the counting modeCounting mode using the Acai function Storing the sample unitPercent mode % 100100 0 % 1000. %Setting example Selecting a comparison modeComparator FuncCp Hi 000000 001234Cp lo 000000 001230Calibration using a weight CalibrationCal 060000Cal 50000 GlpCal 97985 Gravity acceleration correctionEntering the function setting mode FunctionsCp Hi Key operationFunc Poff ACaiEnd Cp Hi Unit End idStoring weighing units Function list Func ACai UminSmpl LdinEJ-03 Installation OptionsEJ-02 USB Interface EJ-03 RS-232C serial interfaceInterface specifications EIA RS-232CData output mode Data formatCommand mode Command listEJ-07 / EJ-08 Installation EJ-07 / EJ-08 UnderhookEJ-11 Breeze break EJ-12 Carrying caseEJ-13 Installation EJ-13 Density Determination KitDensity specific gravity measurement 000 Change the function tableExample of density measurement 2000 1751 8801Setting the ID number ID Number and GLPEnd id func Display character setOutput example Data format for calibration reportData format for calibration test report Calibration test using a weight TitleEnd Block Title block and End block20000 20000 EndMaintenance Error codesOverload error Range over notice Unit weight error Sample quantity noticeError1 Cal e -Cal eLb0 Hb0 Hb1 Lb1Specifications EJ seriesOther weighing units Dimensions OptionsGravity Acceleration MAP Values of gravity at various locationsWorld map