Teledyne 7300A manual Pin Contact, Digital Remote Cal Inputs

Page 19

Infrared Gas Analyzer

Installation 2

 

 

 

 

 

 

 

 

 

System Alarm:

Actuates when DC power supplied to circuits is

 

unacceptable in one or more parameters. Permanently

 

configured as failsafe and latching. Cannot be de-

 

feated. Actuates if self test fails.

 

(Reset by pressing

button to remove power. Then

 

press

again and any other button EXCEPT

System to resume.

Further detail can be found in chapter 4, section 4-5.

Table 2-5: Alarm Relay Contact Pins

Pin Contact

45Threshold Alarm 1, normally closed contact

28Threshold Alarm 1, moving contact

46Threshold Alarm 1, normally open contact

42Threshold Alarm 2, normally closed contact

44Threshold Alarm 2, moving contact

43Threshold Alarm 2, normally open contact

36System Alarm, normally closed contact

20System Alarm, moving contact

37System Alarm, normally open contact

2.3.3.3Digital Remote Cal Inputs

Accept 0 V (off) or 24 V dc (on) inputs for remote control of calibra- tion. (See Remote Calibration Protocol below.) See Table 2-5 for pin connections.

Zero:

Floating input. A 5 to 24 V pulse input across the + and –

 

pins puts the analyzer into the Zero mode. Either side may be

 

grounded at the source of the signal. A synchronous signal

 

must open and close the gas control valves appropriately.

Span:

Floating input. A 5 to 24 V pulse input across the + and –

 

pins puts the analyzer into the Span mode. Either side may

 

be grounded at the source of the signal. A synchronous signal

 

must open and close the gas control valves appropriately.

Cal Contact: This relay contact is closed while analyzer is spanning and/or zeroing. (See Remote Calibration Protocol below.)

Teledyne Analytical Instruments

2-7

Image 19
Contents Teledyne Analytical Instruments Model 7300ACopyright 2000 Teledyne Analytical Instruments Model 7300AApplication Impurity of Interest Infrared Gas Analyzer Technician Record SheetModel 7300A Teledyne Analytical Instruments Model 7300A Installation Table of Contents IntroductionModel 7300A Infrared Gas Analyzer Start-up and Theory of Operation Table of ContentsOperation Electrical/Control Unit Modes/Functioning Appendix MaintenanceTypical Gas Applications OverviewInfrared Gas Analyzer Introduction General Main Features of the AnalyzerIntroduction Model 7300A Ndir Analyzer Introduction Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Installation Installing and Connecting the AnalyzerUnpacking the Analyzer/Inspection Calibration Gases Installation Model 7300AElectrical Power Connections User ConnectionsVenting the System Infrared Gas Analyzer Installation Pipe ConnectionsElectrical Connections Rear Panel Sample Delivery SystemAnalog Outputs Installation Model 7300A Primary Input PowerFuse Installation 3 50-Pin Equipment Interface ConnectorExamples Pin FunctionOutput V dc Output mA dc Range Voltage Current mA Application Alarm RelaysDigital Remote Cal Inputs Pin ContactRemote Calibration Connections Range ID Relays Remote Valve ConnectorNetwork I/O FET Series Resistance 4 RS-232 PortParameter Setting Command DescriptionGas Requirements Calibration Calibration Fluids Testing the SystemCalibration Infrared Gas Analyzer Installation Installation Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Start-up and Operation PreliminaryNdir Analyzer Startup Initial Set-up and ZeroingStart-up and Theory of Operation Model 7300A Operational CalibrationAnalyzer Theory of operation GeneralBlock Diagram of the Model 7300A Electronics Circuit Description Digital Signal Processing & Electronics Linearizer Linearizer Output Linearizer Input Teledyne Analytical Instruments Page Infrared Gas Analyzer Operation IntroductionMode/Function Selection Using the ControlsAnalysis Mode Operation Model 7300ASetup Mode Setup Mode Enter Data EntryEscape Enter Setting up an AUTO-CALEntering the Password Infrared Gas Analyzer Operation Password ProtectionAAA Installing or Changing the PasswordLogging Out System Self-Diagnostic TestChecking Linearity with Algorithm Model ScreenTroubleshooting information Digital Filter Setup Zero Offset Adjustment CAL-OUT Function Adjst output to 4ma ENT to OK Zero Cal Zero and Span FunctionsAuto Mode Zeroing Cell Failure Manual Mode ZeroingSpan Cal Auto Mode SpanningManual Mode Spanning Installation Alarms FunctionPress enter again to set the alarm setpoints Manual Select/Define Range Screen Range Select FunctionAuto Screen Precautions Illustrates these schemes graphically Analyze FunctionSet Application Screen ProgrammingOFFST/INVRT ENT Checking the linearization Curve Algorithm ScreenManual Mode Linearization Auto Mode Linearization Offset OUTPUT/Reversal Output Output Signal Reversal Special Function SetupPolarity Reversal Output Signal OffsetGain Preset Infrared Gas Analyzer Maintenance MaintenanceMaintenance Model 7300A Routine MaintenanceFilter Ndir Analyzer Measurement Cell System Self Diagnostic TestPreamp PowerMajor Internal Components AnalogCell Assembly Troubleshooting General Trouble Shooting Chart Maintenance Model 7300A Infrared Gas Analyzer Maintenance Maintenance Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Appendix Drawing List SpecificationsAppendix Model 7300A Qty Description Recommended 2-Year Spare Parts ListAttachment 7300A Infrared Gas Analyzer Appendix