Teledyne 7300A manual Digital Signal Processing & Electronics

Page 32

3 Start-up and Theory of Operation

Model 7300A

 

 

 

 

 

 

 

 

 

3.5Digital Signal Processing & Electronics

The Model 7300A uses an 8031 microcontroller (Central Processing Unit—CPU) with 32 kB of RAM and 128 kB of ROM to control all signal processing, input/output, and display functions for the analyzer. System power is supplied from a universal power supply module designed (C65507) to be compatible with any international power source. (See Major Internal Components in chapter Maintenance for the location of the power supply and the main electronic PC boards.)

The Temperature Control board (C69535A) is set to a single voltage

(110 or 220 VAC) and set the temperature of the sampling system at 45oC. The signal processing electronics including the microprocessor, analog to digital, and digital to analog converters are located on the Motherboard (C67435B) on side of the case.

The detector output level depends on the intensity of the IR source, the length of the cell and the type of fluid in the cell. Usually the output is between 0.2 and 0.9 mVDC. The Teledyne detector consists of four detec- tors: A, B, C and D. They are strapped to the inputs of the positive, very high-gain amplifiers U1, U2, U3, and U4. Amplifiers U1 through U4 are high quality, very low offset amplifiers.

Preamplified signals (usually between .2 to 1.0 volt) are delivered via the ribbon cable to the measuring board through an 8-point dip switch. Connectors deliver power from the measuring board to the amplifiers and 300–450 mADC current to the IR source.

The motherboard serves as a power distribution and interface board.

The Auto-Zero amplifier PCB board is mounted on top of the Motherboard. These boards are accessible by sliding the system out of the case after removing the screws on the back plate.

The Temperature Control Board keeps the temperature of the measur- ing cell regulated to within 0.1 degree C. A thermistor is used to measure the temperature, and a zero-crossing switch regulates the power of the heaters inside the sample chamber. The result is a sensor output signal that is tem- perature independent.

The output of the preamp is fed to variable gain amplifier, which provides automatic range switching under control of the CPU. The output from the variable gain amplifier is sent to an 18 bit analog to digital con- verter.

The digital concentration signal along with input from the control panel is processed by the CPU and passed on to the 12-bit DAC, which outputs 0-

1V dc Concentration and Range ID signals. An voltage-to-current converter provides 4-20 mA dc concentration signal and range ID outputs.

3-6

Teledyne Analytical Instruments

Image 32
Contents Model 7300A Teledyne Analytical InstrumentsModel 7300A Copyright 2000 Teledyne Analytical InstrumentsInfrared Gas Analyzer Technician Record Sheet Application Impurity of InterestModel 7300A Teledyne Analytical Instruments Model 7300A Model 7300A Infrared Gas Analyzer Table of Contents IntroductionInstallation Operation Electrical/Control Unit Modes/Functioning Table of ContentsStart-up and Theory of Operation Maintenance AppendixInfrared Gas Analyzer Introduction OverviewTypical Gas Applications Introduction Model 7300A Main Features of the AnalyzerGeneral Ndir Analyzer Introduction Model 7300A Teledyne Analytical Instruments Unpacking the Analyzer/Inspection Installing and Connecting the AnalyzerInfrared Gas Analyzer Installation Installation Model 7300A Electrical Power ConnectionsUser Connections Calibration GasesInfrared Gas Analyzer Installation Pipe Connections Electrical Connections Rear PanelSample Delivery System Venting the SystemInstallation Model 7300A Primary Input Power Fuse Installation3 50-Pin Equipment Interface Connector Analog OutputsOutput V dc Output mA dc Pin FunctionExamples Alarm Relays Range Voltage Current mA ApplicationPin Contact Digital Remote Cal InputsRemote Calibration Connections Network I/O Remote Valve ConnectorRange ID Relays 4 RS-232 Port FET Series ResistanceGas Requirements Command DescriptionParameter Setting Calibration Testing the SystemCalibration Calibration Fluids Infrared Gas Analyzer Installation Installation Model 7300A Teledyne Analytical Instruments Preliminary Ndir Analyzer StartupInitial Set-up and Zeroing Infrared Gas Analyzer Start-up and OperationOperational Calibration Start-up and Theory of Operation Model 7300ATheory of operation General AnalyzerBlock Diagram of the Model 7300A Electronics Circuit Description Digital Signal Processing & Electronics Linearizer Linearizer Output Linearizer Input Teledyne Analytical Instruments Page Introduction Infrared Gas Analyzer OperationUsing the Controls Analysis ModeOperation Model 7300A Mode/Function SelectionSetup Mode Setup Mode Escape Data EntryEnter Setting up an AUTO-CAL EnterInfrared Gas Analyzer Operation Password Protection Entering the PasswordInstalling or Changing the Password AAASystem Self-Diagnostic Test Logging OutModel Screen Checking Linearity with AlgorithmTroubleshooting information Digital Filter Setup Zero Offset Adjustment CAL-OUT Function Adjst output to 4ma ENT to OK Zero and Span Functions Zero CalAuto Mode Zeroing Manual Mode Zeroing Cell FailureAuto Mode Spanning Span CalManual Mode Spanning Alarms Function InstallationPress enter again to set the alarm setpoints Range Select Function Manual Select/Define Range ScreenAuto Screen Precautions Analyze Function Illustrates these schemes graphicallyProgramming Set Application ScreenOFFST/INVRT ENT Curve Algorithm Screen Checking the linearizationManual Mode Linearization Auto Mode Linearization Special Function Setup Offset OUTPUT/Reversal Output Output Signal ReversalOutput Signal Offset Polarity ReversalGain Preset Maintenance Infrared Gas Analyzer MaintenanceFilter Routine MaintenanceMaintenance Model 7300A System Self Diagnostic Test Ndir Analyzer Measurement CellPower Major Internal ComponentsAnalog PreampCell Assembly Troubleshooting General Trouble Shooting Chart Maintenance Model 7300A Infrared Gas Analyzer Maintenance Maintenance Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Appendix Appendix Model 7300A SpecificationsDrawing List Recommended 2-Year Spare Parts List Qty DescriptionAttachment 7300A Infrared Gas Analyzer Appendix