Operation 4

 

 

 

 

 

 

 

 

 

is pressed, the Spanreading changes to the correct value. The instrument then automatically enters the Analyze function.

4.5The Alarms Function

The Model 7300A is equipped with 6 fully adjustable set points concentra- tion with two alarms and a system failure alarm relay. Each alarm relay has a set of form “C" contacts rated for 3 amperes resistive load at 250 V ac. See Figure in Chapter 3, Installation and/or the Interconnection Diagram included at the back of this manual for relay terminal connections.

The system failure alarm has a fixed configuration described in chapter 3

Installation.

The concentration alarms can be configured from the front panel as either high or low alarms by the operator. The alarm modes can be set as latching or non-latching, and either fail-safeor non-fail-safe, or, they can be defeated altogether. The setpoints for the alarms are also established using this function.

Decide how your alarms should be configured. The choice will depend upon your process. Consider the following four points:

1.Which if any of the alarms are to be high alarms and which if any are to be low alarms?

Setting an alarm as HIGH triggers the alarm when the contaminant concentration rises above the setpoint. Setting an alarm as LOW triggers the alarm when the contaminant concentration falls below the setpoint.

Decide whether you want the alarms to be set as:

Both high (high and high-high) alarms, or

One high and one low alarm, or

Both low (low and low-low) alarms.

2.Are either or both of the alarms to be configured as failsafe?

In failsafe mode, the alarm relay de-energizes in an alarm condition. For non-failsafe operation, the relay is energized in an alarm condition. You can set either or both of the concentration alarms to operate in failsafe or non-failsafe mode.

3.Are either of the alarms to be latching?

In latching mode, once the alarm or alarms trigger, they will remain in the alarm mode even if process conditions revert back to non-alarm conditions. This mode requires an alarm to be recognized before it can be reset. In the non-latching mode, the

Teledyne Analytical Instruments

4-21

Image 57
Contents Teledyne Analytical Instruments Model 7300ACopyright 2000 Teledyne Analytical Instruments Model 7300AApplication Impurity of Interest Infrared Gas Analyzer Technician Record SheetModel 7300A Teledyne Analytical Instruments Model 7300A Table of Contents Introduction InstallationModel 7300A Infrared Gas Analyzer Table of Contents Start-up and Theory of OperationOperation Electrical/Control Unit Modes/Functioning Appendix MaintenanceOverview Typical Gas ApplicationsInfrared Gas Analyzer Introduction Main Features of the Analyzer GeneralIntroduction Model 7300A Ndir Analyzer Introduction Model 7300A Teledyne Analytical Instruments Installing and Connecting the Analyzer Infrared Gas Analyzer InstallationUnpacking the Analyzer/Inspection Electrical Power Connections Installation Model 7300AUser Connections Calibration GasesElectrical Connections Rear Panel Infrared Gas Analyzer Installation Pipe ConnectionsSample Delivery System Venting the SystemFuse Installation Installation Model 7300A Primary Input Power3 50-Pin Equipment Interface Connector Analog OutputsPin Function ExamplesOutput V dc Output mA dc Range Voltage Current mA Application Alarm RelaysDigital Remote Cal Inputs Pin ContactRemote Calibration Connections Remote Valve Connector Range ID RelaysNetwork I/O FET Series Resistance 4 RS-232 PortCommand Description Parameter SettingGas Requirements Testing the System Calibration Calibration FluidsCalibration Infrared Gas Analyzer Installation Installation Model 7300A Teledyne Analytical Instruments Ndir Analyzer Startup PreliminaryInitial Set-up and Zeroing Infrared Gas Analyzer Start-up and OperationStart-up and Theory of Operation Model 7300A Operational CalibrationAnalyzer Theory of operation GeneralBlock Diagram of the Model 7300A Electronics Circuit Description Digital Signal Processing & Electronics Linearizer Linearizer Output Linearizer Input Teledyne Analytical Instruments Page Infrared Gas Analyzer Operation IntroductionAnalysis Mode Using the ControlsOperation Model 7300A Mode/Function SelectionSetup Mode Setup Mode Data Entry EnterEscape Enter Setting up an AUTO-CALEntering the Password Infrared Gas Analyzer Operation Password ProtectionAAA Installing or Changing the PasswordLogging Out System Self-Diagnostic TestChecking Linearity with Algorithm Model ScreenTroubleshooting information Digital Filter Setup Zero Offset Adjustment CAL-OUT Function Adjst output to 4ma ENT to OK Zero Cal Zero and Span FunctionsAuto Mode Zeroing Cell Failure Manual Mode ZeroingSpan Cal Auto Mode SpanningManual Mode Spanning Installation Alarms FunctionPress enter again to set the alarm setpoints Manual Select/Define Range Screen Range Select FunctionAuto Screen Precautions Illustrates these schemes graphically Analyze FunctionSet Application Screen ProgrammingOFFST/INVRT ENT Checking the linearization Curve Algorithm ScreenManual Mode Linearization Auto Mode Linearization Offset OUTPUT/Reversal Output Output Signal Reversal Special Function SetupPolarity Reversal Output Signal OffsetGain Preset Infrared Gas Analyzer Maintenance MaintenanceRoutine Maintenance Maintenance Model 7300AFilter Ndir Analyzer Measurement Cell System Self Diagnostic TestMajor Internal Components PowerAnalog PreampCell Assembly Troubleshooting General Trouble Shooting Chart Maintenance Model 7300A Infrared Gas Analyzer Maintenance Maintenance Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Appendix Specifications Drawing ListAppendix Model 7300A Qty Description Recommended 2-Year Spare Parts ListAttachment 7300A Infrared Gas Analyzer Appendix