Appendix | Model 7300A | |
|
|
|
|
|
|
|
|
|
ATTACHMENT 7300A
Quote "Exceptions" and "GAS PHASE Conditions" for this application:
1Response Time is proportional to sample system design for
2We assume no particulates that could plug the lines or contaminate the optical components.
3Control the pressure in the measuring sample cell to within 0.5 % of the nominal operating cell pressure (for example: 760 mm ± 4 mm Hg absolute) is required, return pressure assume to be stable at 1 ATM A. Maximum cell operation pressure to be 5 psig based upon stan- dard metal/sapphire cell design.
4The sample flow through the analyzer should be regulated to within ± 5% of its recommended absolute nominal set point flow value. This will avoid any pressure drops across any orifices.
5The temperature of the sample from the
6The sample temperature should not vary at the inlet more than ± 15 OC of the calibration temperature, nor exceed the design operating specifi- cations typically
7Any background components later found to be present and not speci- fied in the original application feasibility may void the instrument perfor- mance should they interfere.
For CO2 the impurity of N2O will interfere. Assume no others at mea- suring wavelength.
8We recommend when ambient temperatures exceed outside a
9Should the general purpose or explosion proof instrument be supplied on a
Teledyne Analytical Instruments |