Teledyne 7300A manual Zero Offset Adjustment

Page 49

Infrared Gas Analyzer

Operation 4

 

 

 

 

 

 

 

 

 

6160

7320

8640

91280

102360

The response time listed above can and will change depending on the application and they merely serve to illustrate the effect of the digital filter. The digital filter disengages if there is an upset that is more than 5% of full scale. As the reading comes within the range, the filter becomes active once again.

4.3.10Zero Offset Adjustment

The software in this instrument provides a way to enter an offset on the zero operation of the analyzer. If the background gas of the process is different than the zero calibration gas being used, the reading will have an offset that will be constant throughout its working range. Thus, the need to provide an offset when the instrument is being zeroed.

How to access the offset function:

To access this function, Scroll through the menu and select OFFSET. Press the Enter key to see the following screen:

Enter zero offset:

0.00 %

The offset value can be modified by using the Up/Down keys. Next section shows how to select this value. Suffice to say that whatever value you enter, will be automatically added to the reading. Thus, if you entered -0.10 %, at the end of the zero the display will show -0.10 %.

When the instrument enters the zero mode in AUTO, the instrument will do the work of bringing the reading back to zero plus the offset value that was entered. If you chose MANual zero mode, then you must adjust the zero of the instrument the corresponding section of the manual but with one difference: instead of bringing the display to read zero, you must make the display read zero plus the value entered as offset.

Teledyne Analytical Instruments

4-13

Image 49
Contents Teledyne Analytical Instruments Model 7300ACopyright 2000 Teledyne Analytical Instruments Model 7300AApplication Impurity of Interest Infrared Gas Analyzer Technician Record SheetModel 7300A Teledyne Analytical Instruments Model 7300A Installation Table of Contents IntroductionModel 7300A Infrared Gas Analyzer Start-up and Theory of Operation Table of ContentsOperation Electrical/Control Unit Modes/Functioning Appendix MaintenanceTypical Gas Applications OverviewInfrared Gas Analyzer Introduction General Main Features of the AnalyzerIntroduction Model 7300A Ndir Analyzer Introduction Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Installation Installing and Connecting the AnalyzerUnpacking the Analyzer/Inspection Electrical Power Connections Installation Model 7300AUser Connections Calibration GasesElectrical Connections Rear Panel Infrared Gas Analyzer Installation Pipe ConnectionsSample Delivery System Venting the SystemFuse Installation Installation Model 7300A Primary Input Power3 50-Pin Equipment Interface Connector Analog OutputsExamples Pin FunctionOutput V dc Output mA dc Range Voltage Current mA Application Alarm RelaysDigital Remote Cal Inputs Pin ContactRemote Calibration Connections Range ID Relays Remote Valve ConnectorNetwork I/O FET Series Resistance 4 RS-232 PortParameter Setting Command DescriptionGas Requirements Calibration Calibration Fluids Testing the SystemCalibration Infrared Gas Analyzer Installation Installation Model 7300A Teledyne Analytical Instruments Ndir Analyzer Startup PreliminaryInitial Set-up and Zeroing Infrared Gas Analyzer Start-up and OperationStart-up and Theory of Operation Model 7300A Operational CalibrationAnalyzer Theory of operation GeneralBlock Diagram of the Model 7300A Electronics Circuit Description Digital Signal Processing & Electronics Linearizer Linearizer Output Linearizer Input Teledyne Analytical Instruments Page Infrared Gas Analyzer Operation IntroductionAnalysis Mode Using the ControlsOperation Model 7300A Mode/Function SelectionSetup Mode Setup Mode Enter Data EntryEscape Enter Setting up an AUTO-CALEntering the Password Infrared Gas Analyzer Operation Password ProtectionAAA Installing or Changing the PasswordLogging Out System Self-Diagnostic TestChecking Linearity with Algorithm Model ScreenTroubleshooting information Digital Filter Setup Zero Offset Adjustment CAL-OUT Function Adjst output to 4ma ENT to OK Zero Cal Zero and Span FunctionsAuto Mode Zeroing Cell Failure Manual Mode ZeroingSpan Cal Auto Mode SpanningManual Mode Spanning Installation Alarms FunctionPress enter again to set the alarm setpoints Manual Select/Define Range Screen Range Select FunctionAuto Screen Precautions Illustrates these schemes graphically Analyze FunctionSet Application Screen ProgrammingOFFST/INVRT ENT Checking the linearization Curve Algorithm ScreenManual Mode Linearization Auto Mode Linearization Offset OUTPUT/Reversal Output Output Signal Reversal Special Function SetupPolarity Reversal Output Signal OffsetGain Preset Infrared Gas Analyzer Maintenance MaintenanceMaintenance Model 7300A Routine MaintenanceFilter Ndir Analyzer Measurement Cell System Self Diagnostic TestMajor Internal Components PowerAnalog PreampCell Assembly Troubleshooting General Trouble Shooting Chart Maintenance Model 7300A Infrared Gas Analyzer Maintenance Maintenance Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Appendix Drawing List SpecificationsAppendix Model 7300A Qty Description Recommended 2-Year Spare Parts ListAttachment 7300A Infrared Gas Analyzer Appendix