Teledyne 7300A manual Circuit Description

Page 31

Infrared Gas Analyzer

Start-up and Operation

 

 

 

 

 

 

 

 

 

3.4 Circuit Description

The Teledyne Analytical Instruments IR bench is a multiple wave- length, single beam design. It uses a quadruple detector that consists of a specially designed, patented thermopile with small IR filters mounted in front of it to produce independent voltages. These correspond to the transmission of energy at each filter’s wavelength. They are amplified by special high stability operational amplifiers to produce the output of the device.

IR energy is generated by a filament (IR source) operating at a moder- ately elevated temperature.

The cell cavity allows the sample gases to pass between the IR source and the detector. Sapphire cell windows allow passage of this IR energy with minimal absorption. The gases that pass through the cell absorb energy of the IR band at different wavelengths within the spectrum.

An optical filter is mounted on the face of each detector. Each detector responds to a particular wavelength, producing a voltage proportional to the energy that has passed through the gas and filter. If the gas concentration changes, the energy at the measuring wavelength changes, changing only the measuring voltage output. The reference voltage output is not affected by the gas concentration changes. This “measurement” voltage change is electroni- cally processed as the instrument’s output signal

IR Bench Block Diagram

Teledyne Analytical Instruments

3-5

Image 31
Contents Teledyne Analytical Instruments Model 7300ACopyright 2000 Teledyne Analytical Instruments Model 7300AApplication Impurity of Interest Infrared Gas Analyzer Technician Record SheetModel 7300A Teledyne Analytical Instruments Model 7300A Installation Table of Contents IntroductionModel 7300A Infrared Gas Analyzer Start-up and Theory of Operation Table of ContentsOperation Electrical/Control Unit Modes/Functioning Appendix MaintenanceTypical Gas Applications OverviewInfrared Gas Analyzer Introduction General Main Features of the AnalyzerIntroduction Model 7300A Ndir Analyzer Introduction Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Installation Installing and Connecting the AnalyzerUnpacking the Analyzer/Inspection Calibration Gases Installation Model 7300AElectrical Power Connections User ConnectionsVenting the System Infrared Gas Analyzer Installation Pipe ConnectionsElectrical Connections Rear Panel Sample Delivery SystemAnalog Outputs Installation Model 7300A Primary Input PowerFuse Installation 3 50-Pin Equipment Interface ConnectorExamples Pin FunctionOutput V dc Output mA dc Range Voltage Current mA Application Alarm RelaysDigital Remote Cal Inputs Pin ContactRemote Calibration Connections Range ID Relays Remote Valve ConnectorNetwork I/O FET Series Resistance 4 RS-232 PortParameter Setting Command DescriptionGas Requirements Calibration Calibration Fluids Testing the SystemCalibration Infrared Gas Analyzer Installation Installation Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Start-up and Operation PreliminaryNdir Analyzer Startup Initial Set-up and ZeroingStart-up and Theory of Operation Model 7300A Operational CalibrationAnalyzer Theory of operation GeneralBlock Diagram of the Model 7300A Electronics Circuit Description Digital Signal Processing & Electronics Linearizer Linearizer Output Linearizer Input Teledyne Analytical Instruments Page Infrared Gas Analyzer Operation IntroductionMode/Function Selection Using the ControlsAnalysis Mode Operation Model 7300ASetup Mode Setup Mode Enter Data EntryEscape Enter Setting up an AUTO-CALEntering the Password Infrared Gas Analyzer Operation Password ProtectionAAA Installing or Changing the PasswordLogging Out System Self-Diagnostic TestChecking Linearity with Algorithm Model ScreenTroubleshooting information Digital Filter Setup Zero Offset Adjustment CAL-OUT Function Adjst output to 4ma ENT to OK Zero Cal Zero and Span FunctionsAuto Mode Zeroing Cell Failure Manual Mode ZeroingSpan Cal Auto Mode SpanningManual Mode Spanning Installation Alarms FunctionPress enter again to set the alarm setpoints Manual Select/Define Range Screen Range Select FunctionAuto Screen Precautions Illustrates these schemes graphically Analyze FunctionSet Application Screen ProgrammingOFFST/INVRT ENT Checking the linearization Curve Algorithm ScreenManual Mode Linearization Auto Mode Linearization Offset OUTPUT/Reversal Output Output Signal Reversal Special Function SetupPolarity Reversal Output Signal OffsetGain Preset Infrared Gas Analyzer Maintenance MaintenanceMaintenance Model 7300A Routine MaintenanceFilter Ndir Analyzer Measurement Cell System Self Diagnostic TestPreamp PowerMajor Internal Components AnalogCell Assembly Troubleshooting General Trouble Shooting Chart Maintenance Model 7300A Infrared Gas Analyzer Maintenance Maintenance Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Appendix Drawing List SpecificationsAppendix Model 7300A Qty Description Recommended 2-Year Spare Parts ListAttachment 7300A Infrared Gas Analyzer Appendix