Infrared Gas Analyzer | Operation 4 | |
|
|
|
|
|
|
|
|
|
If the concentration detected is over range, the first line of the display blinks continuously.
4.8Programming
CAUTION: The programming functions of the Set Range and Curve Algorithm screens are configured at the facto- ry to the users application specification. These func- tions should only be reprogrammed by trained, qualified personnel.
To program, you must:
1.Enter the password, if you are using the analyzer’s password protection capability.
2.Connect a computer or computer terminal capable of sending an
Now you will be able to select the APPLICATION and ALGORITHM setupfunctions.
4.8.1 The Set Application Screen
The Set Application screen allows reprogramming of the three analysis ranges and the calibration range (including impurity gas, background gas, low end of range, high end of range, and % or ppm units). Original programming is usually done at the factory according to the customer’s application. It must be done through the
Note: It is important to distinguish between this System program- ming subfunction and the Range button function, which is an operator control. The Set Range Screen of the Application function allows the user to DEFINE the upper and lower limits of a range AND the application of the range. The Range func- tion only allows the user to select or define the limits, or to select the application, but not to define the application.
Normally the Model 7300A is factory set to default to manual range selection, unless it is ordered as a
Teledyne Analytical Instruments |