Teledyne 7300A manual Preliminary, Ndir Analyzer Startup, Initial Set-up and Zeroing

Page 27

Infrared Gas Analyzer

Start-up and Operation

 

 

 

 

 

 

 

 

 

3.0Start-up and Operation

3.1Preliminary

Before applying power to the system, TAI suggests that the electrical wiring installation be checked against the system input-output diagram. Proper attention to this preliminary check will prevent severe damage caused by wiring errors.

Also, verify that all connections to the system have been made correctly. Refer to the system outline diagram for proper connections.

3.2 NDIR Analyzer Startup

Before power is supplied to the analyzer, TAI recommends the follow- ing operator’s check be performed:

1.Check for loose or damaged components.

2.Verify that all plug-in circuit cards are firmly seated in their receptacles.

3.Inspect and verify that all wiring connections are in agreement with the system Interconnection diagram.

4.Check for correct span gases. (See 4.2: Preparation for Calibra- tion).

5.Check that sample pressure is regulated to ± .5% of nominal operating pressure.

6.Check and assure sample remains above dewpoint to eliminate any condensation from sample tap to return point.

Power up the unit by depressing the rear panel switch. From a first time power-on warm start attempt, allow (1) one hour warm-up to proceed. Observe that the Digital display will go through a diagnostic routine before the readings revert to a continuous concentration readout.

3.2.1 Initial Set-up and Zeroing

Assure the sample will enter from the zero inlet gas position. Open the zero gas tank and set the pressure regulator to 20 psig. Set the zero gas flow

Teledyne Analytical Instruments

3-1

Image 27
Contents Teledyne Analytical Instruments Model 7300ACopyright 2000 Teledyne Analytical Instruments Model 7300AApplication Impurity of Interest Infrared Gas Analyzer Technician Record SheetModel 7300A Teledyne Analytical Instruments Model 7300A Table of Contents Introduction InstallationModel 7300A Infrared Gas Analyzer Table of Contents Start-up and Theory of OperationOperation Electrical/Control Unit Modes/Functioning Appendix MaintenanceOverview Typical Gas ApplicationsInfrared Gas Analyzer Introduction Main Features of the Analyzer GeneralIntroduction Model 7300A Ndir Analyzer Introduction Model 7300A Teledyne Analytical Instruments Installing and Connecting the Analyzer Infrared Gas Analyzer InstallationUnpacking the Analyzer/Inspection Calibration Gases Installation Model 7300AElectrical Power Connections User ConnectionsVenting the System Infrared Gas Analyzer Installation Pipe ConnectionsElectrical Connections Rear Panel Sample Delivery SystemAnalog Outputs Installation Model 7300A Primary Input PowerFuse Installation 3 50-Pin Equipment Interface ConnectorPin Function ExamplesOutput V dc Output mA dc Range Voltage Current mA Application Alarm RelaysDigital Remote Cal Inputs Pin ContactRemote Calibration Connections Remote Valve Connector Range ID RelaysNetwork I/O FET Series Resistance 4 RS-232 PortCommand Description Parameter SettingGas Requirements Testing the System Calibration Calibration FluidsCalibration Infrared Gas Analyzer Installation Installation Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Start-up and Operation PreliminaryNdir Analyzer Startup Initial Set-up and ZeroingStart-up and Theory of Operation Model 7300A Operational CalibrationAnalyzer Theory of operation GeneralBlock Diagram of the Model 7300A Electronics Circuit Description Digital Signal Processing & Electronics Linearizer Linearizer Output Linearizer Input Teledyne Analytical Instruments Page Infrared Gas Analyzer Operation IntroductionMode/Function Selection Using the ControlsAnalysis Mode Operation Model 7300ASetup Mode Setup Mode Data Entry EnterEscape Enter Setting up an AUTO-CALEntering the Password Infrared Gas Analyzer Operation Password ProtectionAAA Installing or Changing the PasswordLogging Out System Self-Diagnostic TestChecking Linearity with Algorithm Model ScreenTroubleshooting information Digital Filter Setup Zero Offset Adjustment CAL-OUT Function Adjst output to 4ma ENT to OK Zero Cal Zero and Span FunctionsAuto Mode Zeroing Cell Failure Manual Mode ZeroingSpan Cal Auto Mode SpanningManual Mode Spanning Installation Alarms FunctionPress enter again to set the alarm setpoints Manual Select/Define Range Screen Range Select FunctionAuto Screen Precautions Illustrates these schemes graphically Analyze FunctionSet Application Screen ProgrammingOFFST/INVRT ENT Checking the linearization Curve Algorithm ScreenManual Mode Linearization Auto Mode Linearization Offset OUTPUT/Reversal Output Output Signal Reversal Special Function SetupPolarity Reversal Output Signal OffsetGain Preset Infrared Gas Analyzer Maintenance MaintenanceRoutine Maintenance Maintenance Model 7300AFilter Ndir Analyzer Measurement Cell System Self Diagnostic TestPreamp PowerMajor Internal Components AnalogCell Assembly Troubleshooting General Trouble Shooting Chart Maintenance Model 7300A Infrared Gas Analyzer Maintenance Maintenance Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Appendix Specifications Drawing ListAppendix Model 7300A Qty Description Recommended 2-Year Spare Parts ListAttachment 7300A Infrared Gas Analyzer Appendix