Infrared Gas Analyzer | Operation 4 | |
|
|
|
|
|
|
|
|
|
If you want to calibrate ALL of the ranges at once (multiple application analyzers only), use auto mode zeroing in the Cal Range.
Make sure the zero gas is flowing to the instrument. If you get a CELL CANNOT BE BALANCED message while zeroing skip to section 4.4.1.3.
4.4.1.1Auto Mode Zeroing
Observe the precautions in sections 4.4 and 4.4.1, above.Scroll to ZERO function buy using UP/DOWN control and enter the zero function mode. The screen allows you to select whether the zero calibration is to be performed automatically or manually. Use the UP/DOWN key to toggle between AUTO and MAN zero settling. Stop when AUTO appears, blinking, on the display.
Select zero mode: AUTO
Press Enter to begin zeroing.
####.## % CO2 Slope=#.###
The beginning zero level is shown in the upper left corner of the display. As the zero reading settles, the screen displays and updates information on Slope= in ppm/second (unless the Slope starts within the acceptable zero range and does not need to settle further). The system first does a course zero, shown in the lower right corner of the screen as
Then, and whenever Slope is less than 0.01 for at least 3 min, instead of Slope you will see a countdown: 9 Left, 8 Left, and so fourth. These are software steps in the zeroing process that the system must complete, AFTER settling, before it can go back to Analyze. Software zero is indicated by
####.## % | CO2 |
4 Left=#.### |
The zeroing process will automatically conclude when the output is within the acceptable range for a good zero. Then the analyzer automatically returns to the Analyze mode.
Teledyne Analytical Instruments |