Operation 4

 

 

 

 

 

 

 

 

 

range readings are accurate UNLESS the application uses linearization over the selected range.

The concentration ranges can be redefined using the Range function Manual screen, and the application gases can be redefined using the APPLICA- TION function, if they are not already defined as necessary.

CAUTION: Redefining applications or ranges might require re- linearization and/or recalibration.

Tosetupautomaticranging:

Select range on the MAIN MENU, and ENTER to start the Range func-

tion.

Select range mode : AUTO

Note: If all three ranges are currently defined for different applica- tion gases, then the above screen does not display (because mode must be manual).

If above screen displays, use the UP/DOWN key to Select AUTO, and Enter.

Press Escape to return to the previous Analyze Function.

4.6.3 Precautions

The Model 7300A allows a great deal of flexibility in choosing ranges for automatic range switching. However, there are some pitfalls that are to be avoided.

Ranges that work well together are:

Ranges that have the same lower limits but upper limits that differ by approximately an order of magnitude

Ranges whose upper limits coincide with the lower limits of the next higher range

Ranges where there is a gap between the upper limit of the range and the lower limit of the next higher range.

Range schemes that are to be avoided include:

Teledyne Analytical Instruments

4-25

Image 61
Contents Teledyne Analytical Instruments Model 7300ACopyright 2000 Teledyne Analytical Instruments Model 7300AApplication Impurity of Interest Infrared Gas Analyzer Technician Record SheetModel 7300A Teledyne Analytical Instruments Model 7300A Installation Table of Contents IntroductionModel 7300A Infrared Gas Analyzer Start-up and Theory of Operation Table of ContentsOperation Electrical/Control Unit Modes/Functioning Appendix MaintenanceTypical Gas Applications OverviewInfrared Gas Analyzer Introduction General Main Features of the AnalyzerIntroduction Model 7300A Ndir Analyzer Introduction Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Installation Installing and Connecting the AnalyzerUnpacking the Analyzer/Inspection Electrical Power Connections Installation Model 7300AUser Connections Calibration GasesElectrical Connections Rear Panel Infrared Gas Analyzer Installation Pipe ConnectionsSample Delivery System Venting the SystemFuse Installation Installation Model 7300A Primary Input Power3 50-Pin Equipment Interface Connector Analog OutputsExamples Pin FunctionOutput V dc Output mA dc Range Voltage Current mA Application Alarm RelaysDigital Remote Cal Inputs Pin ContactRemote Calibration Connections Range ID Relays Remote Valve ConnectorNetwork I/O FET Series Resistance 4 RS-232 PortParameter Setting Command DescriptionGas Requirements Calibration Calibration Fluids Testing the SystemCalibration Infrared Gas Analyzer Installation Installation Model 7300A Teledyne Analytical Instruments Ndir Analyzer Startup PreliminaryInitial Set-up and Zeroing Infrared Gas Analyzer Start-up and OperationStart-up and Theory of Operation Model 7300A Operational CalibrationAnalyzer Theory of operation GeneralBlock Diagram of the Model 7300A Electronics Circuit Description Digital Signal Processing & Electronics Linearizer Linearizer Output Linearizer Input Teledyne Analytical Instruments Page Infrared Gas Analyzer Operation IntroductionAnalysis Mode Using the ControlsOperation Model 7300A Mode/Function SelectionSetup Mode Setup Mode Enter Data EntryEscape Enter Setting up an AUTO-CALEntering the Password Infrared Gas Analyzer Operation Password ProtectionAAA Installing or Changing the PasswordLogging Out System Self-Diagnostic TestChecking Linearity with Algorithm Model ScreenTroubleshooting information Digital Filter Setup Zero Offset Adjustment CAL-OUT Function Adjst output to 4ma ENT to OK Zero Cal Zero and Span FunctionsAuto Mode Zeroing Cell Failure Manual Mode ZeroingSpan Cal Auto Mode SpanningManual Mode Spanning Installation Alarms FunctionPress enter again to set the alarm setpoints Manual Select/Define Range Screen Range Select FunctionAuto Screen Precautions Illustrates these schemes graphically Analyze FunctionSet Application Screen ProgrammingOFFST/INVRT ENT Checking the linearization Curve Algorithm ScreenManual Mode Linearization Auto Mode Linearization Offset OUTPUT/Reversal Output Output Signal Reversal Special Function SetupPolarity Reversal Output Signal OffsetGain Preset Infrared Gas Analyzer Maintenance MaintenanceMaintenance Model 7300A Routine MaintenanceFilter Ndir Analyzer Measurement Cell System Self Diagnostic TestMajor Internal Components PowerAnalog PreampCell Assembly Troubleshooting General Trouble Shooting Chart Maintenance Model 7300A Infrared Gas Analyzer Maintenance Maintenance Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Appendix Drawing List SpecificationsAppendix Model 7300A Qty Description Recommended 2-Year Spare Parts ListAttachment 7300A Infrared Gas Analyzer Appendix