4 Operation

Model 7300A

 

 

 

 

 

 

 

 

 

In the autoranging mode, the microprocessor automatically responds to concentration changes by switching ranges for optimum readout sensitivity. If the upper limit of the operating range is reached, the instrument automatically shifts to the next higher range. If the concentration falls to below 85% of full scale of the next lower range, the instrument switches to the lower range. A correspond- ing shift in the DC concentration output, and in the range ID outputs, will be noticed.

The autoranging feature can be overridden so that analog output stays on a fixed range regardless of the contaminant concentration detected. If the concen- tration exceeds the upper limit of the range, the DC output will saturate at 1 V dc (20 mA at the current output).

However, the digital readout and the RS-232 output of the concentration are unaffected by the fixed range. They continue to read beyond the full-scale setting until amplifier saturation is reached. Below amplifier saturation, the over range readings are accurate UNLESS the application uses linearization over the selected range.

To program the ranges, you must first perform the four steps indicated at the beginning of section 4.8 Programming. You will then be in the MAIN MENU and selecting application function screen.

Sel rng to set appl: —> Ø1 Ø2 Ø3 CAL <—

Use the UP/DOWN key to increment/decrement the range number to 01, 02, 03, or CAL, and Enter.

Imp: CO2

FR:Ø TO:1Ø %

Use the UP/DOWN key to increment the respective parameters as de- sired.

Use the ENTER to move from Imp: (impurity) to ppm or % to FR: (from- lower end of ranges), and TO: (to upper-end ranges).

Last Enter will accept the values and the screen will display

OFFST/INVRT: <ENT>

Standard: <ESC>

ESC if your application is standard. If your application requires OFFSET or Reversal output, ENTER will set your output as OFFSET or REVERSAL. (See special function setup section 4.9 for more information).

. (See note below.) Repeat for each range you want to set.

4-28

Teledyne Analytical Instruments

Image 64
Contents Model 7300A Teledyne Analytical InstrumentsModel 7300A Copyright 2000 Teledyne Analytical InstrumentsInfrared Gas Analyzer Technician Record Sheet Application Impurity of InterestModel 7300A Teledyne Analytical Instruments Model 7300A Installation Table of Contents IntroductionModel 7300A Infrared Gas Analyzer Start-up and Theory of Operation Table of ContentsOperation Electrical/Control Unit Modes/Functioning Maintenance AppendixTypical Gas Applications OverviewInfrared Gas Analyzer Introduction General Main Features of the AnalyzerIntroduction Model 7300A Ndir Analyzer Introduction Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Installation Installing and Connecting the AnalyzerUnpacking the Analyzer/Inspection Installation Model 7300A Electrical Power ConnectionsUser Connections Calibration GasesInfrared Gas Analyzer Installation Pipe Connections Electrical Connections Rear PanelSample Delivery System Venting the SystemInstallation Model 7300A Primary Input Power Fuse Installation3 50-Pin Equipment Interface Connector Analog OutputsExamples Pin FunctionOutput V dc Output mA dc Alarm Relays Range Voltage Current mA ApplicationPin Contact Digital Remote Cal InputsRemote Calibration Connections Range ID Relays Remote Valve ConnectorNetwork I/O 4 RS-232 Port FET Series ResistanceParameter Setting Command DescriptionGas Requirements Calibration Calibration Fluids Testing the SystemCalibration Infrared Gas Analyzer Installation Installation Model 7300A Teledyne Analytical Instruments Preliminary Ndir Analyzer StartupInitial Set-up and Zeroing Infrared Gas Analyzer Start-up and OperationOperational Calibration Start-up and Theory of Operation Model 7300ATheory of operation General AnalyzerBlock Diagram of the Model 7300A Electronics Circuit Description Digital Signal Processing & Electronics Linearizer Linearizer Output Linearizer Input Teledyne Analytical Instruments Page Introduction Infrared Gas Analyzer OperationUsing the Controls Analysis ModeOperation Model 7300A Mode/Function SelectionSetup Mode Setup Mode Enter Data EntryEscape Setting up an AUTO-CAL EnterInfrared Gas Analyzer Operation Password Protection Entering the PasswordInstalling or Changing the Password AAASystem Self-Diagnostic Test Logging OutModel Screen Checking Linearity with AlgorithmTroubleshooting information Digital Filter Setup Zero Offset Adjustment CAL-OUT Function Adjst output to 4ma ENT to OK Zero and Span Functions Zero CalAuto Mode Zeroing Manual Mode Zeroing Cell FailureAuto Mode Spanning Span CalManual Mode Spanning Alarms Function InstallationPress enter again to set the alarm setpoints Range Select Function Manual Select/Define Range ScreenAuto Screen Precautions Analyze Function Illustrates these schemes graphicallyProgramming Set Application ScreenOFFST/INVRT ENT Curve Algorithm Screen Checking the linearizationManual Mode Linearization Auto Mode Linearization Special Function Setup Offset OUTPUT/Reversal Output Output Signal ReversalOutput Signal Offset Polarity ReversalGain Preset Maintenance Infrared Gas Analyzer MaintenanceMaintenance Model 7300A Routine MaintenanceFilter System Self Diagnostic Test Ndir Analyzer Measurement CellPower Major Internal ComponentsAnalog PreampCell Assembly Troubleshooting General Trouble Shooting Chart Maintenance Model 7300A Infrared Gas Analyzer Maintenance Maintenance Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Appendix Drawing List SpecificationsAppendix Model 7300A Recommended 2-Year Spare Parts List Qty DescriptionAttachment 7300A Infrared Gas Analyzer Appendix