Teledyne 7300A manual Remote Calibration Connections

Page 20

2 Installation

Model 7300A

 

 

 

 

 

 

 

 

 

Table 2-6: Remote Calibration Connections

Pin Function

9+ Remote Zero

11– Remote Zero

10+ Remote Span

12– Remote Span

40Cal Contact

41Cal Contact

Remote Calibration Protocol: To properly time the Digital Remote Cal Inputs to the Model 7300A Analyzer, the customer's controller must monitor the Cal Relay Contact.

When the contact is OPEN, the analyzer is analyzing, the Remote Cal Inputs are being polled, and a zero or span command can be sent.

When the contact is CLOSED, the analyzer is already calibrating. It will ignore your request to calibrate, and it will not remember that request.

Once a zero or span command is sent, and acknowledged (contact closes), release it. If the command is continued until after the zero or span is complete, the calibration will repeat and the Cal Relay Contact (CRC) will close again.

For example:

1)Test the CRC. When the CRC is open, Send a zero command until the CRC closes (The CRC will close quickly.)

2)When the CRC closes, remove the zero command.

3)When CRC opens again, send a span command until the CRC closes. (The CRC will close quickly.)

4)When the CRC closes, remove the span command.

When CRC opens again, zero and span are done, and the sample is being analyzed.

Note: The Remote Probe connector provides signals to operate the zero and span gas valves synchronously. However, if you have the –C Internal valve option, which includes zero and span gas inputs, the 7300A automatically regulates the zero, span and sample gas flow.

2-8

Teledyne Analytical Instruments

Image 20
Contents Model 7300A Teledyne Analytical InstrumentsModel 7300A Copyright 2000 Teledyne Analytical InstrumentsInfrared Gas Analyzer Technician Record Sheet Application Impurity of InterestModel 7300A Teledyne Analytical Instruments Model 7300A Model 7300A Infrared Gas Analyzer Table of Contents IntroductionInstallation Operation Electrical/Control Unit Modes/Functioning Table of ContentsStart-up and Theory of Operation Maintenance AppendixInfrared Gas Analyzer Introduction OverviewTypical Gas Applications Introduction Model 7300A Main Features of the AnalyzerGeneral Ndir Analyzer Introduction Model 7300A Teledyne Analytical Instruments Unpacking the Analyzer/Inspection Installing and Connecting the AnalyzerInfrared Gas Analyzer Installation Installation Model 7300A Electrical Power ConnectionsUser Connections Calibration GasesInfrared Gas Analyzer Installation Pipe Connections Electrical Connections Rear PanelSample Delivery System Venting the SystemInstallation Model 7300A Primary Input Power Fuse Installation3 50-Pin Equipment Interface Connector Analog OutputsOutput V dc Output mA dc Pin FunctionExamples Alarm Relays Range Voltage Current mA ApplicationPin Contact Digital Remote Cal InputsRemote Calibration Connections Network I/O Remote Valve ConnectorRange ID Relays 4 RS-232 Port FET Series ResistanceGas Requirements Command DescriptionParameter Setting Calibration Testing the SystemCalibration Calibration Fluids Infrared Gas Analyzer Installation Installation Model 7300A Teledyne Analytical Instruments Preliminary Ndir Analyzer StartupInitial Set-up and Zeroing Infrared Gas Analyzer Start-up and OperationOperational Calibration Start-up and Theory of Operation Model 7300ATheory of operation General AnalyzerBlock Diagram of the Model 7300A Electronics Circuit Description Digital Signal Processing & Electronics Linearizer Linearizer Output Linearizer Input Teledyne Analytical Instruments Page Introduction Infrared Gas Analyzer OperationUsing the Controls Analysis ModeOperation Model 7300A Mode/Function SelectionSetup Mode Setup Mode Escape Data EntryEnter Setting up an AUTO-CAL EnterInfrared Gas Analyzer Operation Password Protection Entering the PasswordInstalling or Changing the Password AAASystem Self-Diagnostic Test Logging OutModel Screen Checking Linearity with AlgorithmTroubleshooting information Digital Filter Setup Zero Offset Adjustment CAL-OUT Function Adjst output to 4ma ENT to OK Zero and Span Functions Zero CalAuto Mode Zeroing Manual Mode Zeroing Cell FailureAuto Mode Spanning Span CalManual Mode Spanning Alarms Function InstallationPress enter again to set the alarm setpoints Range Select Function Manual Select/Define Range ScreenAuto Screen Precautions Analyze Function Illustrates these schemes graphicallyProgramming Set Application ScreenOFFST/INVRT ENT Curve Algorithm Screen Checking the linearizationManual Mode Linearization Auto Mode Linearization Special Function Setup Offset OUTPUT/Reversal Output Output Signal ReversalOutput Signal Offset Polarity ReversalGain Preset Maintenance Infrared Gas Analyzer MaintenanceFilter Routine MaintenanceMaintenance Model 7300A System Self Diagnostic Test Ndir Analyzer Measurement CellPower Major Internal ComponentsAnalog PreampCell Assembly Troubleshooting General Trouble Shooting Chart Maintenance Model 7300A Infrared Gas Analyzer Maintenance Maintenance Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Appendix Appendix Model 7300A SpecificationsDrawing List Recommended 2-Year Spare Parts List Qty DescriptionAttachment 7300A Infrared Gas Analyzer Appendix