Infrared Gas Analyzer

Maintenance 5

 

 

 

 

 

 

 

 

 

5.3 NDIR Analyzer Measurement Cell

The Infrared Analyzer contains an auto zero circuit which automatically zeroes the analyzer. This zeroing compensates for dirt in the sample cell and, under normal circumstances, will compensate for very large amounts of contamination in the cell. If there is a need for removing the cell and cleaning it, follow this procedure:

A)Turn off power to the analyzer and open the optical bench housing. Disconnect the cell heater wires (if present).

B)Remove the cell assembly from the optical bench. Non-sealed cells can be swabbed with water and alcohol and dried thoroughly before reassembly. Sealed cells can be blown out with air, washed, or replaced.

C)After the cells have been cleaned and thoroughly dried,

reassemble them into the optical bench. Also attach any disconnected wiring back onto the bench and check all connections and fasteners.

D)Turn on the analyzer and allow it to warmÄup for at least ten minutes. Check for zero and span gas response.

NOTE: Typically, sample cell should only be removed for cleaning when the Auto Zero circuit can no longer com~ensate for offset or if there is a loss of sensitivity caused by cell lnefficiency .

5.4System Self Diagnostic Test

1.Press the System button to enter the system mode.

2.Use the < > arrow keys to move to More, and press Enter.

3.Use the < > arrow keys to move to Self-Test, and press Enter. The following failure codes apply:

Teledyne Analytical Instruments

5-3

Image 73
Contents Teledyne Analytical Instruments Model 7300ACopyright 2000 Teledyne Analytical Instruments Model 7300AApplication Impurity of Interest Infrared Gas Analyzer Technician Record SheetModel 7300A Teledyne Analytical Instruments Model 7300A Installation Table of Contents IntroductionModel 7300A Infrared Gas Analyzer Start-up and Theory of Operation Table of ContentsOperation Electrical/Control Unit Modes/Functioning Appendix MaintenanceTypical Gas Applications OverviewInfrared Gas Analyzer Introduction General Main Features of the AnalyzerIntroduction Model 7300A Ndir Analyzer Introduction Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Installation Installing and Connecting the AnalyzerUnpacking the Analyzer/Inspection Electrical Power Connections Installation Model 7300AUser Connections Calibration GasesElectrical Connections Rear Panel Infrared Gas Analyzer Installation Pipe ConnectionsSample Delivery System Venting the SystemFuse Installation Installation Model 7300A Primary Input Power3 50-Pin Equipment Interface Connector Analog OutputsExamples Pin FunctionOutput V dc Output mA dc Range Voltage Current mA Application Alarm RelaysDigital Remote Cal Inputs Pin ContactRemote Calibration Connections Range ID Relays Remote Valve ConnectorNetwork I/O FET Series Resistance 4 RS-232 PortParameter Setting Command DescriptionGas Requirements Calibration Calibration Fluids Testing the SystemCalibration Infrared Gas Analyzer Installation Installation Model 7300A Teledyne Analytical Instruments Ndir Analyzer Startup PreliminaryInitial Set-up and Zeroing Infrared Gas Analyzer Start-up and OperationStart-up and Theory of Operation Model 7300A Operational CalibrationAnalyzer Theory of operation GeneralBlock Diagram of the Model 7300A Electronics Circuit Description Digital Signal Processing & Electronics Linearizer Linearizer Output Linearizer Input Teledyne Analytical Instruments Page Infrared Gas Analyzer Operation IntroductionAnalysis Mode Using the ControlsOperation Model 7300A Mode/Function SelectionSetup Mode Setup Mode Enter Data EntryEscape Enter Setting up an AUTO-CALEntering the Password Infrared Gas Analyzer Operation Password ProtectionAAA Installing or Changing the PasswordLogging Out System Self-Diagnostic TestChecking Linearity with Algorithm Model ScreenTroubleshooting information Digital Filter Setup Zero Offset Adjustment CAL-OUT Function Adjst output to 4ma ENT to OK Zero Cal Zero and Span FunctionsAuto Mode Zeroing Cell Failure Manual Mode ZeroingSpan Cal Auto Mode SpanningManual Mode Spanning Installation Alarms FunctionPress enter again to set the alarm setpoints Manual Select/Define Range Screen Range Select FunctionAuto Screen Precautions Illustrates these schemes graphically Analyze FunctionSet Application Screen ProgrammingOFFST/INVRT ENT Checking the linearization Curve Algorithm ScreenManual Mode Linearization Auto Mode Linearization Offset OUTPUT/Reversal Output Output Signal Reversal Special Function SetupPolarity Reversal Output Signal OffsetGain Preset Infrared Gas Analyzer Maintenance MaintenanceMaintenance Model 7300A Routine MaintenanceFilter Ndir Analyzer Measurement Cell System Self Diagnostic TestMajor Internal Components PowerAnalog PreampCell Assembly Troubleshooting General Trouble Shooting Chart Maintenance Model 7300A Infrared Gas Analyzer Maintenance Maintenance Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Appendix Drawing List SpecificationsAppendix Model 7300A Qty Description Recommended 2-Year Spare Parts ListAttachment 7300A Infrared Gas Analyzer Appendix