4 Operation

Model 7300A

 

 

 

 

 

 

 

 

 

If you do not wish to continue a function, you can abort the session by escaping . Escaping a function takes the analyzer back to the previous screen, or to the ANALYZE Function, depending on the function escaped.

reproduced, at the appropriate point in the procedure, in a Mono- spaced type style. Push-button names are printed in Oblique type.

4.3.2 Setting up an AUTO-CAL

When proper automatic valving is connected (see chapter 3, installation), the Analyzer can cycle itself through a sequence of steps that automatically zero and span the instrument.

Note: Before setting up an AUTO-CAL, be sure you understand the Zero and Span functions as described in section 4.4, and follow the precautions given there.

Note: If you require highly accurate AUTO-CAL timing, use external AUTO-CAL control where possible. The internal clock in the Model 7300A is accurate to 2-3 %. Accordingly, internally scheduled calibrations can vary 2-3 % per day.

Note: If your ranges are configured for different applications, then AUTO-CAL will calibrate all of the ranges simultaneously (by calibrating the Cal Range).

To setup an AUTO-CALcycle:

TheVFDwilldisplayfivesubfunctions.

Call out MAIN MENU, scroll to AUTO-CAL function, and ENTER. A new screen for ZERO/SPAN set appears.

ZERO

in

Ød

Øh

off

SPAN

in

Ød

Øh

off

Use UP or DOWN key to blink ZERO (or SPAN), then Enter. (You

won’t be able to set OFF to ON if a zero interval is entered.) A Span Every ...

(or Zero Every ...) screen appears.

Zero schedule: OFF

Day: Ød Hour: Øh

Use UP or DOWN key to set a value in days, then ENTER to move to the start-time value in hours. Use UP or DOWN keys to set a start-time value, then

ENTER.

To turn ON the SPAN and/or ZERO cycles (to activate AUTO–CAL): useUP or DOWN keys to set the OFF/ON field to ON. You can now turn these fields ON because there is a nonzero span time defined.

4-6

Teledyne Analytical Instruments

Image 42
Contents Model 7300A Teledyne Analytical InstrumentsModel 7300A Copyright 2000 Teledyne Analytical InstrumentsInfrared Gas Analyzer Technician Record Sheet Application Impurity of InterestModel 7300A Teledyne Analytical Instruments Model 7300A Table of Contents Introduction InstallationModel 7300A Infrared Gas Analyzer Table of Contents Start-up and Theory of OperationOperation Electrical/Control Unit Modes/Functioning Maintenance AppendixOverview Typical Gas ApplicationsInfrared Gas Analyzer Introduction Main Features of the Analyzer GeneralIntroduction Model 7300A Ndir Analyzer Introduction Model 7300A Teledyne Analytical Instruments Installing and Connecting the Analyzer Infrared Gas Analyzer InstallationUnpacking the Analyzer/Inspection User Connections Installation Model 7300AElectrical Power Connections Calibration GasesSample Delivery System Infrared Gas Analyzer Installation Pipe ConnectionsElectrical Connections Rear Panel Venting the System3 50-Pin Equipment Interface Connector Installation Model 7300A Primary Input PowerFuse Installation Analog OutputsPin Function ExamplesOutput V dc Output mA dc Alarm Relays Range Voltage Current mA ApplicationPin Contact Digital Remote Cal InputsRemote Calibration Connections Remote Valve Connector Range ID RelaysNetwork I/O 4 RS-232 Port FET Series ResistanceCommand Description Parameter SettingGas Requirements Testing the System Calibration Calibration FluidsCalibration Infrared Gas Analyzer Installation Installation Model 7300A Teledyne Analytical Instruments Initial Set-up and Zeroing PreliminaryNdir Analyzer Startup Infrared Gas Analyzer Start-up and OperationOperational Calibration Start-up and Theory of Operation Model 7300ATheory of operation General AnalyzerBlock Diagram of the Model 7300A Electronics Circuit Description Digital Signal Processing & Electronics Linearizer Linearizer Output Linearizer Input Teledyne Analytical Instruments Page Introduction Infrared Gas Analyzer OperationOperation Model 7300A Using the ControlsAnalysis Mode Mode/Function SelectionSetup Mode Setup Mode Data Entry EnterEscape Setting up an AUTO-CAL EnterInfrared Gas Analyzer Operation Password Protection Entering the PasswordInstalling or Changing the Password AAASystem Self-Diagnostic Test Logging OutModel Screen Checking Linearity with AlgorithmTroubleshooting information Digital Filter Setup Zero Offset Adjustment CAL-OUT Function Adjst output to 4ma ENT to OK Zero and Span Functions Zero CalAuto Mode Zeroing Manual Mode Zeroing Cell FailureAuto Mode Spanning Span CalManual Mode Spanning Alarms Function InstallationPress enter again to set the alarm setpoints Range Select Function Manual Select/Define Range ScreenAuto Screen Precautions Analyze Function Illustrates these schemes graphicallyProgramming Set Application ScreenOFFST/INVRT ENT Curve Algorithm Screen Checking the linearizationManual Mode Linearization Auto Mode Linearization Special Function Setup Offset OUTPUT/Reversal Output Output Signal ReversalOutput Signal Offset Polarity ReversalGain Preset Maintenance Infrared Gas Analyzer MaintenanceRoutine Maintenance Maintenance Model 7300AFilter System Self Diagnostic Test Ndir Analyzer Measurement CellAnalog PowerMajor Internal Components PreampCell Assembly Troubleshooting General Trouble Shooting Chart Maintenance Model 7300A Infrared Gas Analyzer Maintenance Maintenance Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Appendix Specifications Drawing ListAppendix Model 7300A Recommended 2-Year Spare Parts List Qty DescriptionAttachment 7300A Infrared Gas Analyzer Appendix