4 Operation | Model 7300A | |
|
|
|
|
|
|
|
|
|
Note: The span gas use to span the analyzer must be >90% of the range being analyzed.
Before starting linearization, perform a standard calibration. See section 4.4. To enter data:
From the MAIN MENU screen—
1.Use UP/DOWN to select ALGORITHM , and Enter.
2.Select and Enter SETUP.
3.Enter AUTO from the Calibration Mode Select screen. The Auto Linearize Mode data entry screen appears.
1.95 % CO2 Input(Ø) :2.00
5.Use the UP/DOWN keys to set the proper value of calibration gas, and Enter. Repeat this step for each
6.Repeat step 5 for each of the special calibration gases, from the lowest to the highest concentrations. Escape when done.
To end the session, send:
st<enter> st<enter>
to the analyzer from the computer.
4.9 Special Function Setup
CAUTION: The programming functions of the output signal reversal, polarity reversal and gain preset are config- ured at the factory to the users application specifica- tion. These functions should only be reprogrammed by trained, qualified personnel.
4.9.1 OFFSET OUTPUT/Reversal Output
4.9.1.1 Output Signal Reversal
Some applications require a reversal of the output signals in order for the
Teledyne Analytical Instruments |