4 Operation

Model 7300A

 

 

 

 

 

 

 

 

 

Note: The span gas use to span the analyzer must be >90% of the range being analyzed.

Before starting linearization, perform a standard calibration. See section 4.4. To enter data:

From the MAIN MENU screen—

1.Use UP/DOWN to select ALGORITHM , and Enter.

2.Select and Enter SETUP.

3.Enter AUTO from the Calibration Mode Select screen. The Auto Linearize Mode data entry screen appears.

1.95 % CO2 Input(Ø) :2.00

5.Use the UP/DOWN keys to set the proper value of calibration gas, and Enter. Repeat this step for each cal-point number as it appears in the Input (x) parentheses.

6.Repeat step 5 for each of the special calibration gases, from the lowest to the highest concentrations. Escape when done.

To end the session, send:

st<enter> st<enter>

to the analyzer from the computer.

4.9 Special Function Setup

CAUTION: The programming functions of the output signal reversal, polarity reversal and gain preset are config- ured at the factory to the users application specifica- tion. These functions should only be reprogrammed by trained, qualified personnel.

4.9.1 OFFSET OUTPUT/Reversal Output

4.9.1.1 Output Signal Reversal

Some applications require a reversal of the output signals in order for the 4-20mA and 0-1 V DC output signals to correspond with the low and high end of the concentration range. For example, if an application involves the analysis of 85-100% oxygen in a background of argon by measuring the thermal conduc-

4-32

Teledyne Analytical Instruments

Image 68
Contents Model 7300A Teledyne Analytical InstrumentsModel 7300A Copyright 2000 Teledyne Analytical InstrumentsInfrared Gas Analyzer Technician Record Sheet Application Impurity of InterestModel 7300A Teledyne Analytical Instruments Model 7300A Model 7300A Infrared Gas Analyzer Table of Contents IntroductionInstallation Operation Electrical/Control Unit Modes/Functioning Table of ContentsStart-up and Theory of Operation Maintenance AppendixInfrared Gas Analyzer Introduction OverviewTypical Gas Applications Introduction Model 7300A Main Features of the AnalyzerGeneral Ndir Analyzer Introduction Model 7300A Teledyne Analytical Instruments Unpacking the Analyzer/Inspection Installing and Connecting the AnalyzerInfrared Gas Analyzer Installation Installation Model 7300A Electrical Power ConnectionsUser Connections Calibration GasesInfrared Gas Analyzer Installation Pipe Connections Electrical Connections Rear PanelSample Delivery System Venting the SystemInstallation Model 7300A Primary Input Power Fuse Installation3 50-Pin Equipment Interface Connector Analog OutputsOutput V dc Output mA dc Pin FunctionExamples Alarm Relays Range Voltage Current mA ApplicationPin Contact Digital Remote Cal InputsRemote Calibration Connections Network I/O Remote Valve ConnectorRange ID Relays 4 RS-232 Port FET Series ResistanceGas Requirements Command DescriptionParameter Setting Calibration Testing the SystemCalibration Calibration Fluids Infrared Gas Analyzer Installation Installation Model 7300A Teledyne Analytical Instruments Preliminary Ndir Analyzer StartupInitial Set-up and Zeroing Infrared Gas Analyzer Start-up and OperationOperational Calibration Start-up and Theory of Operation Model 7300ATheory of operation General AnalyzerBlock Diagram of the Model 7300A Electronics Circuit Description Digital Signal Processing & Electronics Linearizer Linearizer Output Linearizer Input Teledyne Analytical Instruments Page Introduction Infrared Gas Analyzer OperationUsing the Controls Analysis ModeOperation Model 7300A Mode/Function SelectionSetup Mode Setup Mode Escape Data EntryEnter Setting up an AUTO-CAL EnterInfrared Gas Analyzer Operation Password Protection Entering the PasswordInstalling or Changing the Password AAASystem Self-Diagnostic Test Logging OutModel Screen Checking Linearity with AlgorithmTroubleshooting information Digital Filter Setup Zero Offset Adjustment CAL-OUT Function Adjst output to 4ma ENT to OK Zero and Span Functions Zero CalAuto Mode Zeroing Manual Mode Zeroing Cell FailureAuto Mode Spanning Span CalManual Mode Spanning Alarms Function InstallationPress enter again to set the alarm setpoints Range Select Function Manual Select/Define Range ScreenAuto Screen Precautions Analyze Function Illustrates these schemes graphicallyProgramming Set Application ScreenOFFST/INVRT ENT Curve Algorithm Screen Checking the linearizationManual Mode Linearization Auto Mode Linearization Special Function Setup Offset OUTPUT/Reversal Output Output Signal ReversalOutput Signal Offset Polarity ReversalGain Preset Maintenance Infrared Gas Analyzer MaintenanceFilter Routine MaintenanceMaintenance Model 7300A System Self Diagnostic Test Ndir Analyzer Measurement CellPower Major Internal ComponentsAnalog PreampCell Assembly Troubleshooting General Trouble Shooting Chart Maintenance Model 7300A Infrared Gas Analyzer Maintenance Maintenance Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Appendix Appendix Model 7300A SpecificationsDrawing List Recommended 2-Year Spare Parts List Qty DescriptionAttachment 7300A Infrared Gas Analyzer Appendix