4 Operation | Model 7300A | |
|
|
|
|
|
|
|
|
|
4.4.1.2Manual Mode Zeroing
Scroll to Zero and enter the Zero function. The screen that appears allows you to select between automatic or manual zero calibration. Use the UP/DOWN keys to toggle between AUTO and MAN zero settling. Stop when MANUAL appears, blinking, on the display.
Select zero mode: MANUAL
Enter to begin the zero calibration. After a few seconds the first of three zeroing screens appears. The number in the upper left hand corner is the first- stage zero offset. The microprocessor samples the output at a predetermined rate.
####.## | % | CO2 |
Zero adj:2048 |
The analyzer goes through
####.## % CO2 Slope=#.###
Generally, you have a good zero when Slope is less than 0.05 ppm/s for about 30 seconds.
Once zero settling completes, the information is stored in the analyzer’s memory, and the instrument automatically returns to the Analyze mode.
4.4.1.3 Cell Failure
Cell failure in the 7300A is usually associated with inability to zero the instrument with a reasonable voltage differential across the Wheatstone bridge. If this should ever happen, the 7300A system alarm trips, and the VFD displays a failuremessage.
Cell cannot be balanced
Check your zero gas
Before replacing the sensor:
a. Check your zero gas to make sure it is within specifications.
Teledyne Analytical Instruments |