Teledyne 7300A manual Digital Filter Setup

Page 48

4 Operation

Model 7300A

 

 

 

 

 

 

 

 

 

Hard_offset_C=3015

(This is the raw DAC count of the Coarse zero

 

adjustment. It should read between 0 and 4095)

Hard_offset_F=2715

(This is the raw DAC count of the Fine zero

 

adjustment. It should read between 0 and 4095)

Current_gain = 4

(This is the gain the analyzer is on, it should read

 

between 0 and 9)

4.3.9Digital Filter Setup

The 7300A analyzer has the option of decreasing or increasing the amount of filtering on the signal. This feature enhances the basic filtering done by the analog circuits by setting the amount of digital filtering effected by the micro- processor. To access the digital filter setup, scroll through the menu and select FILTER. Press the Enter key to see the following screen:

Enter Digital

Filter: 4

The number on the second row can be adjusted from 0, minimum digital filtering, to 10, maximum digital filtering. The default setting is 4 and that should suffice for most applications. In some applications where speeding the response time with some trade off in noise is of value, the operator could decrease the number of the digital filter. In applications where the signal is noisy, the operator could switch to a higher number; the response time is slowed down though.

10 to 90% response time on the different settings to a step input is shown below. This response time does not include the contributions of the bench sampling system and the preamplifier near the detector.

Setting

10-90% Response

 

(seconds)

0

2.5

1

5

2

10

3

20

4

40

5

80

 

 

4-12

Teledyne Analytical Instruments

Image 48
Contents Model 7300A Teledyne Analytical InstrumentsModel 7300A Copyright 2000 Teledyne Analytical InstrumentsInfrared Gas Analyzer Technician Record Sheet Application Impurity of InterestModel 7300A Teledyne Analytical Instruments Model 7300A Table of Contents Introduction InstallationModel 7300A Infrared Gas Analyzer Table of Contents Start-up and Theory of OperationOperation Electrical/Control Unit Modes/Functioning Maintenance AppendixOverview Typical Gas ApplicationsInfrared Gas Analyzer Introduction Main Features of the Analyzer GeneralIntroduction Model 7300A Ndir Analyzer Introduction Model 7300A Teledyne Analytical Instruments Installing and Connecting the Analyzer Infrared Gas Analyzer InstallationUnpacking the Analyzer/Inspection Installation Model 7300A Electrical Power ConnectionsUser Connections Calibration GasesInfrared Gas Analyzer Installation Pipe Connections Electrical Connections Rear PanelSample Delivery System Venting the SystemInstallation Model 7300A Primary Input Power Fuse Installation3 50-Pin Equipment Interface Connector Analog OutputsPin Function ExamplesOutput V dc Output mA dc Alarm Relays Range Voltage Current mA ApplicationPin Contact Digital Remote Cal InputsRemote Calibration Connections Remote Valve Connector Range ID RelaysNetwork I/O 4 RS-232 Port FET Series ResistanceCommand Description Parameter SettingGas Requirements Testing the System Calibration Calibration FluidsCalibration Infrared Gas Analyzer Installation Installation Model 7300A Teledyne Analytical Instruments Preliminary Ndir Analyzer StartupInitial Set-up and Zeroing Infrared Gas Analyzer Start-up and OperationOperational Calibration Start-up and Theory of Operation Model 7300ATheory of operation General AnalyzerBlock Diagram of the Model 7300A Electronics Circuit Description Digital Signal Processing & Electronics Linearizer Linearizer Output Linearizer Input Teledyne Analytical Instruments Page Introduction Infrared Gas Analyzer OperationUsing the Controls Analysis ModeOperation Model 7300A Mode/Function SelectionSetup Mode Setup Mode Data Entry EnterEscape Setting up an AUTO-CAL EnterInfrared Gas Analyzer Operation Password Protection Entering the PasswordInstalling or Changing the Password AAASystem Self-Diagnostic Test Logging OutModel Screen Checking Linearity with AlgorithmTroubleshooting information Digital Filter Setup Zero Offset Adjustment CAL-OUT Function Adjst output to 4ma ENT to OK Zero and Span Functions Zero CalAuto Mode Zeroing Manual Mode Zeroing Cell FailureAuto Mode Spanning Span CalManual Mode Spanning Alarms Function InstallationPress enter again to set the alarm setpoints Range Select Function Manual Select/Define Range ScreenAuto Screen Precautions Analyze Function Illustrates these schemes graphicallyProgramming Set Application ScreenOFFST/INVRT ENT Curve Algorithm Screen Checking the linearizationManual Mode Linearization Auto Mode Linearization Special Function Setup Offset OUTPUT/Reversal Output Output Signal ReversalOutput Signal Offset Polarity ReversalGain Preset Maintenance Infrared Gas Analyzer MaintenanceRoutine Maintenance Maintenance Model 7300AFilter System Self Diagnostic Test Ndir Analyzer Measurement CellPower Major Internal ComponentsAnalog PreampCell Assembly Troubleshooting General Trouble Shooting Chart Maintenance Model 7300A Infrared Gas Analyzer Maintenance Maintenance Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Appendix Specifications Drawing ListAppendix Model 7300A Recommended 2-Year Spare Parts List Qty DescriptionAttachment 7300A Infrared Gas Analyzer Appendix