Teledyne 7300A manual Output Signal Offset, Polarity Reversal

Page 69

Infrared Gas Analyzer

Operation 4

 

 

 

 

 

 

 

 

 

tivity of the binary gas, the analyzer would normally be set up so that the 100% oxygen (0% argon) concentration would correspond to the zero level (4mA 0 V) of the output signal. Then, 85% oxygen (15% argon) would correspond to 20mA (1 V) in the signal output.

It may be convenient for the user to have the outputs reversed so that the 85-100% oxygen level outputs a 4-20mA (0-1 V) signal respectively. This can be accomplished by reversing the data input to the custom settings. Not all applications will require a reversing function, however, if this is desirable. This can be programmed by allowing the user to set the analyzer to read reversal output. Contact the factory for further information.

4.9.1.2Output Signal Offset

TAI has provided the output offset feature in the software for the accuracy purpose. In many cases, the analyzer does not require this feature. For exmple, if the analyzer has setup to analyze the sample gas of 40-50% CO2 in nitrogen, normally zero gas of this application requires 40% CO2 in nitrogen. However, 100% of nitrogen can be used to zero the analyzer. In this case, the output offset is not needed to setup. For linear output the accuracy will access success- fully within +/-1% off. But, if the application is analyzing the sample gas that is not linearly, the accuracy of the analyzer may not meet the specification. There- fore, output offset is required and 40% CO2 in nitrogen is also needed for zero calibrationgas.

To set up the output reversal or output offset, see section 4.8.1-set AP- PLICATION sceen.

NOTE: If the inverting has been setup, “i” shall display on the left bottom corner. Otherwise, the left bottom corner display ”n”.

NOTE: In the example 40% CO2 in N2 is discouraged, because it is hard to produce the exact concentration.

4.9.2 Polarity Reversal

In some special applications, user will find that the display dispalys negative concentration values, even if proper span gas is injected. To correct this problem, TAI has added the Polarity Correction feature. This feature can be set asfollow:

Close S1-5 range 1

Close S1-6 range 2

Close S1-7 range 3

Close S1-8 cal range

Teledyne Analytical Instruments

4-33

Image 69
Contents Teledyne Analytical Instruments Model 7300ACopyright 2000 Teledyne Analytical Instruments Model 7300AApplication Impurity of Interest Infrared Gas Analyzer Technician Record SheetModel 7300A Teledyne Analytical Instruments Model 7300A Table of Contents Introduction InstallationModel 7300A Infrared Gas Analyzer Table of Contents Start-up and Theory of OperationOperation Electrical/Control Unit Modes/Functioning Appendix MaintenanceOverview Typical Gas ApplicationsInfrared Gas Analyzer Introduction Main Features of the Analyzer GeneralIntroduction Model 7300A Ndir Analyzer Introduction Model 7300A Teledyne Analytical Instruments Installing and Connecting the Analyzer Infrared Gas Analyzer InstallationUnpacking the Analyzer/Inspection Electrical Power Connections Installation Model 7300AUser Connections Calibration GasesElectrical Connections Rear Panel Infrared Gas Analyzer Installation Pipe ConnectionsSample Delivery System Venting the SystemFuse Installation Installation Model 7300A Primary Input Power3 50-Pin Equipment Interface Connector Analog OutputsPin Function ExamplesOutput V dc Output mA dc Range Voltage Current mA Application Alarm RelaysDigital Remote Cal Inputs Pin ContactRemote Calibration Connections Remote Valve Connector Range ID RelaysNetwork I/O FET Series Resistance 4 RS-232 PortCommand Description Parameter SettingGas Requirements Testing the System Calibration Calibration FluidsCalibration Infrared Gas Analyzer Installation Installation Model 7300A Teledyne Analytical Instruments Ndir Analyzer Startup PreliminaryInitial Set-up and Zeroing Infrared Gas Analyzer Start-up and OperationStart-up and Theory of Operation Model 7300A Operational CalibrationAnalyzer Theory of operation GeneralBlock Diagram of the Model 7300A Electronics Circuit Description Digital Signal Processing & Electronics Linearizer Linearizer Output Linearizer Input Teledyne Analytical Instruments Page Infrared Gas Analyzer Operation IntroductionAnalysis Mode Using the ControlsOperation Model 7300A Mode/Function SelectionSetup Mode Setup Mode Data Entry EnterEscape Enter Setting up an AUTO-CALEntering the Password Infrared Gas Analyzer Operation Password ProtectionAAA Installing or Changing the PasswordLogging Out System Self-Diagnostic TestChecking Linearity with Algorithm Model ScreenTroubleshooting information Digital Filter Setup Zero Offset Adjustment CAL-OUT Function Adjst output to 4ma ENT to OK Zero Cal Zero and Span FunctionsAuto Mode Zeroing Cell Failure Manual Mode ZeroingSpan Cal Auto Mode SpanningManual Mode Spanning Installation Alarms FunctionPress enter again to set the alarm setpoints Manual Select/Define Range Screen Range Select FunctionAuto Screen Precautions Illustrates these schemes graphically Analyze FunctionSet Application Screen ProgrammingOFFST/INVRT ENT Checking the linearization Curve Algorithm ScreenManual Mode Linearization Auto Mode Linearization Offset OUTPUT/Reversal Output Output Signal Reversal Special Function SetupPolarity Reversal Output Signal OffsetGain Preset Infrared Gas Analyzer Maintenance MaintenanceRoutine Maintenance Maintenance Model 7300AFilter Ndir Analyzer Measurement Cell System Self Diagnostic TestMajor Internal Components PowerAnalog PreampCell Assembly Troubleshooting General Trouble Shooting Chart Maintenance Model 7300A Infrared Gas Analyzer Maintenance Maintenance Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Appendix Specifications Drawing ListAppendix Model 7300A Qty Description Recommended 2-Year Spare Parts ListAttachment 7300A Infrared Gas Analyzer Appendix