Teledyne 7300A manual Press enter again to set the alarm setpoints

Page 58

4 Operation

Model 7300A

 

 

 

 

 

 

 

 

 

alarm status will terminate when process conditions revert to non- alarm conditions.

4.Are either of the alarms to be defeated?

The defeat alarm mode is incorporated into the alarm circuit so that maintenance can be performed under conditions which would normally activate the alarms.

The defeat function can also be used to reset a latched alarm. (See procedures, below.)

If you are using password protection, you will need to enter your password to access the alarm functions. Follow the instructions in section 4.3.3 to enter your password. Once you have clearance to proceed, enter the Alarm function.

Note: If all ranges are for the same application, set any one of them will automatically set the others.

Press the Alarm button on the front panel to enter the Alarm function. Make sure that 01 is blinking.

Sel rng to set alm:

—> Ø1 Ø2 Ø3

<—

Set up the Range 1 alarm by moving the blinking over to 01 using the UP/ DOWN arrow keys. Then Enter. Check the gas application and range limits as displayed on the screen.

Gas use: CO2

Range: 0 — 10 %

Press enter again to set the alarm setpoints.

Sel %/ppm alm to set

AL1—PPM AL2—PPM

Use the UP/DOWN keys to choose between % or ppm units. Then Enter to move to the next screen.

AL1: 1ØØØ ppm HI

Dft:N Fs:N Ltch:N

Five parameters can be changed on this screen:

Value of the alarm setpoint, AL1: ####

Out-of-range direction, HI or LO

Defeated? Dft:Y/N (Yes/No)

Failsafe? Fs:Y/N (Yes/No)

Latching? Ltch:Y/N (Yes/No).

To define the setpoint, use the UP/DOWN key while screen is blinking over to AL1: ####. Use the UP/DOWN key to change

4-22

Teledyne Analytical Instruments

Image 58
Contents Model 7300A Teledyne Analytical InstrumentsModel 7300A Copyright 2000 Teledyne Analytical InstrumentsInfrared Gas Analyzer Technician Record Sheet Application Impurity of InterestModel 7300A Teledyne Analytical Instruments Model 7300A Installation Table of Contents IntroductionModel 7300A Infrared Gas Analyzer Start-up and Theory of Operation Table of ContentsOperation Electrical/Control Unit Modes/Functioning Maintenance AppendixTypical Gas Applications OverviewInfrared Gas Analyzer Introduction General Main Features of the AnalyzerIntroduction Model 7300A Ndir Analyzer Introduction Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Installation Installing and Connecting the AnalyzerUnpacking the Analyzer/Inspection User Connections Installation Model 7300AElectrical Power Connections Calibration GasesSample Delivery System Infrared Gas Analyzer Installation Pipe ConnectionsElectrical Connections Rear Panel Venting the System3 50-Pin Equipment Interface Connector Installation Model 7300A Primary Input PowerFuse Installation Analog OutputsExamples Pin FunctionOutput V dc Output mA dc Alarm Relays Range Voltage Current mA ApplicationPin Contact Digital Remote Cal InputsRemote Calibration Connections Range ID Relays Remote Valve ConnectorNetwork I/O 4 RS-232 Port FET Series ResistanceParameter Setting Command DescriptionGas Requirements Calibration Calibration Fluids Testing the SystemCalibration Infrared Gas Analyzer Installation Installation Model 7300A Teledyne Analytical Instruments Initial Set-up and Zeroing PreliminaryNdir Analyzer Startup Infrared Gas Analyzer Start-up and OperationOperational Calibration Start-up and Theory of Operation Model 7300ATheory of operation General AnalyzerBlock Diagram of the Model 7300A Electronics Circuit Description Digital Signal Processing & Electronics Linearizer Linearizer Output Linearizer Input Teledyne Analytical Instruments Page Introduction Infrared Gas Analyzer OperationOperation Model 7300A Using the ControlsAnalysis Mode Mode/Function SelectionSetup Mode Setup Mode Enter Data EntryEscape Setting up an AUTO-CAL EnterInfrared Gas Analyzer Operation Password Protection Entering the PasswordInstalling or Changing the Password AAASystem Self-Diagnostic Test Logging OutModel Screen Checking Linearity with AlgorithmTroubleshooting information Digital Filter Setup Zero Offset Adjustment CAL-OUT Function Adjst output to 4ma ENT to OK Zero and Span Functions Zero CalAuto Mode Zeroing Manual Mode Zeroing Cell FailureAuto Mode Spanning Span CalManual Mode Spanning Alarms Function InstallationPress enter again to set the alarm setpoints Range Select Function Manual Select/Define Range ScreenAuto Screen Precautions Analyze Function Illustrates these schemes graphicallyProgramming Set Application ScreenOFFST/INVRT ENT Curve Algorithm Screen Checking the linearizationManual Mode Linearization Auto Mode Linearization Special Function Setup Offset OUTPUT/Reversal Output Output Signal ReversalOutput Signal Offset Polarity ReversalGain Preset Maintenance Infrared Gas Analyzer MaintenanceMaintenance Model 7300A Routine MaintenanceFilter System Self Diagnostic Test Ndir Analyzer Measurement CellAnalog PowerMajor Internal Components PreampCell Assembly Troubleshooting General Trouble Shooting Chart Maintenance Model 7300A Infrared Gas Analyzer Maintenance Maintenance Model 7300A Teledyne Analytical Instruments Infrared Gas Analyzer Appendix Drawing List SpecificationsAppendix Model 7300A Recommended 2-Year Spare Parts List Qty DescriptionAttachment 7300A Infrared Gas Analyzer Appendix