-399 to -300: Device-Specific Errors

ESG Family Signal Generators

Out of memory; Histogram display cannot function.

There is insufficient working memory available. Report this error to the nearest Agilent Technologies sales or service office.

Out of memory; Impulse response display cannot function.

There is insufficient working memory available. Report this error to the nearest Agilent Technologies sales or service office.

Out of memory; There is insufficient memory to copy the current pattern. The secondary pattern is unchanged.

There was insufficient room to copy or create the current data generation pattern data in order to make it a secondary pattern. To avoid this, set up the smaller of the two patterns first.

Out of memory; There is insufficient memory to store the secondary frame in Pattern RAM. The secondary frame has been turned off.

There was insufficient room to copy or create the current data generation pattern data in order to make it a secondary pattern. To avoid this, set up the smaller of the two patterns first.

Out of memory; Insufficient RAM

A memory comparison between a shadow RAM data area and the corresponding EEPROM data area could not be performed due to insufficient working RAM. This does not necessarily imply a memory problem, since this comparison is only used in stringent diagnostic situations.

Out of memory; I/Q map display cannot function.

There is insufficient working memory available. Report this error to the nearest Agilent Technologies sales or service office.

Out of memory; List formation

The device was unable to allocate space for a lookup table, such as for list mode precalculation. List mode cannot run until either fewer frequencies have been supplied or more memory becomes available and the same set of frequencies are sent again, FREQ:MODE CW is executed, or :FREQ:MODE LIST is executed.

10

Error Messages

Page 12
Image 12