ESG Family Signal Generators

-399 to -300: Device-Specific Errors

 

Self-test failed; Burst Generator Memory Test @ 0x____

 

The burst generator memory failed. Modulation data produced by the

 

burst generator may not be correct. However, if an Unable to check

 

Burst Generator Memory error was also seen, this result is not

 

conclusive. The address of the first location that failed is reported. Report

 

this error to the nearest Agilent Technologies sales or service office.

 

Self-test failed; Chips ___, ___ aliased @ 0x____

 

The data generator memory failed. Modulation data produced by the data

 

generator may not be correct. An address that appeared to be aliased

 

across multiple memory chips has been reported. Report this error to the

 

nearest Agilent Technologies sales or service office.

 

Self-test failed; Data Generator Memory Test @ 0x____

 

The data generator memory failed. Modulation data produced by the data

 

generator may not be correct. However, if an Unable to check Data

 

Generator Memory error was also seen, this result is not conclusive. The

 

address of the first location that failed is reported. Report this error to the

 

nearest Agilent Technologies sales or service office.

 

Self-test failed; EEPROM header checksum error <card_name>.

 

The card identification header for a hardware card is incorrect. If the card

 

is not properly identified, the instrument is likely to be non-functional.

 

Report this error to the nearest Agilent Technologies sales or service office.

 

Self-test failed; Power supply self-test failure

 

The self-test for a particular power supply voltage has failed. The

 

instrument is likely not functional. Report this error to the nearest

 

Agilent Technologies sales or service office.

-321

Out of memory

 

An internal operation needed more memory than was available.

Out of memory

There is insufficient working memory available. Report this error to the nearest Agilent Technologies sales or service office.

Out of memory

Insufficient working memory during arbitrary waveform generation. Report this error to the nearest Agilent Technologies sales or service office.

Error Messages

7

Page 9
Image 9
Agilent Technologies E4400-90333 manual Burst generator memory failed. Modulation data produced by, 321