-299 to -200: Execution Errors

ESG Family Signal Generators

-220

Parameter error

 

A program data element related error has occurred. This particular error

 

message is used if the device cannot detect a more specific error described

 

for errors -221 through -229.

 

Parameter error; <modulation file name>

 

The specified file is not a modulation file type.

-215

Arm deadlock

 

The arm source for the initiation of a measurement is set to GET and a

 

subsequent measurement query is received. The measurement cannot

 

begin until a GET is received, but the GET would cause an INTERRUPTED

 

error.

-214

Trigger deadlock

 

A trigger source for the initiation of a measurement is set to GET and a

 

subsequent measurement query is received. The measurement cannot

 

begin until a GET is received, but the GET would cause an INTERRUPTED

 

error.

 

Trigger deadlock

 

TRIGger:SOURce BUS was selected when a MEASurement? query was

 

executed. No data is returned.

-213

Init ignored

 

A request for a measurement initiation was ignored as another

 

measurement was already in progress.

Init ignored; Cannot initiate sweep in manual mode.

The manual mode is on and therefore the instrument cannot sweep.

Init ignored; Sweep is already initiated.

The list/sweep is currently initiated and sweeping, therefore the command is not legal according to SCPI.

Init ignored; Sweep is already continuously initiated.

The list/sweep is continuously initiated and sweeping, therefore the command is not legal according to SCPI.

46

Error Messages

Page 48
Image 48
Agilent Technologies E4400-90333 220, For errors -221 through, Specified file is not a modulation file type, 215, Error, 214