ESG Family Signal Generators

Introduction

If an error condition occurs in the signal generator, it is reported to both the front panel display error queue and the SCPI (remote interface) error queue. These two queues are viewed and managed separately.

NOTE

When there is an unviewed message in the front panel error queue, the ERR

 

annunciator appears on the signal generator’s display.

 

 

 

 

 

 

 

 

 

Characteristic

Front Panel Display

SCPI Remote Interface

 

 

Error Queue

Error Queue

 

 

 

Capacity (#errors)

30

30

 

 

 

 

 

 

Circular (rotating).

Linear, first-in/first-out.

Overflow Handling

Drops oldest error as new error comes in.

Replaces newest error

 

with: -350, Queue

 

 

 

 

 

 

overflow

 

 

 

 

 

 

Press: Utility > Error Info > View Next

Use SCPI query

Viewing Entries

SYSTem:ERRor? or

(or Previous) Error Message

 

 

STATus:QUEue?

 

 

 

 

 

 

 

 

 

 

Power up

Clearing the Queue

Press: Utility > Error Info > Clear Error Queue(s)

Send a *CLS command

 

 

 

Read last item in the queue

 

 

 

Permanent Errors

 

 

(errors that must be resolved.

Re-reported after queue is cleared.

Re-reported after queue is

For example: unlock,

cleared.

 

avalanched, and hi/lo)

 

 

 

 

 

 

1

Page 3
Image 3
Agilent Technologies E4400-90333 manual Introduction