-299 to -200: Execution Errors

ESG Family Signal Generators

Hardware missing; no such SIO address

The expected board is not installed. Report this error to the nearest

Agilent Technologies sales or service office.

-240Hardware error

A legal program command or query could not be executed because of a hardware problem in the device. The definition of what constitutes a hardware problem is completely device-specific. This error is used when the device cannot detect the more specific errors described for errors -241 through -249.

-233Invalid version

A legal program data element was parsed but could not be executed because the version of the data is incorrect to the device. This particular error is used when file or block data elements are recognized by the instrument, but cannot be executed for reasons of version incompatibility. For example, a non-supported file version or a non-supported instrument version.

-232Invalid format

A legal program data element was parsed but could not be executed because the data format or structure is inappropriate. For example, when loading memory tables or when sending a SYSTem:SET parameter for an unknown instrument.

-231Data questionable

The measurement accuracy is questionable.

Data questionable; EEPROM copy of <filename>.

The EEPROM copy of a file has a correctable error. The system automatically performs the correction. A potential cause is a failing EEPROM. Report this error to the nearest Agilent Technologies sales or service office.

Data questionable; RAM copy of <filename>.

The non-volatile RAM copy of a file has a correctable error. The system automatically performs the correction. A potential cause is a failing backup battery.

-230Data corrupt or stale

Possibly invalid data. A new reading was started but not completed since last access.