-399 to -300: Device-Specific Errors

ESG Family Signal Generators

Configuration memory lost; Persistent state checksum is bad. Using factory defaults.

The persistent state is corrupt and had to be reinitialized with the factory preset values.

Configuration memory lost; Persistent state preset. Using factory defaults.

The persistent state has been forced to return to factory preset values.

 

Configuration memory lost; Persistent state version is bad.

 

Using factory defaults.

 

The persistent state version is not recognized as valid and is assumed to

 

be corrupt. The persistent state is reinitialized with the factory preset

 

values.

-314

Save/recall memory loss

 

The non-volatile data saved by the *SAV? command has been lost.

-313

Calibration memory lost

 

Indicates that non-volatile calibration data has been lost.

-312

PUD memory lost

 

The protected user data saved by the *PUD command has been lost.

-311

Memory error

 

An error was detected in the device’s memory.

 

Memory error; Unable to configure Save Recall registers from

 

non-volatile memory. Save Recall registers re-initialized.

 

Indicates that saved states are no longer usable. Delete explicitly using

 

Catalog.

-310

System error; Absolute I and Q voltages must be at most

 

1 volt.

An internal error has occurred that results in the normalized I or Q voltage to be greater than 1 volt. Report this error to the nearest Agilent Technologies sales or service office.

14

Error Messages

Page 16
Image 16
Agilent Technologies E4400-90333 Values, 314, Non-volatile data saved by the *SAV? command has been lost, 313, 312, 311