-299 to -200: Execution Errors

ESG Family Signal Generators

 

Data corrupt or stale; License file is obsolete, missing

 

entries defaulted.

 

This error can only occur during the instrument’s first power up after

 

having downloaded new firmware. It indicates that the software license

 

file is missing entries for options supported by the current revision of

 

firmware. The missing entries have been added to the file, and their

 

license key values have been defaulted to zero. This error will occurs

 

during the first power up after downloading new firmware into the

 

instrument.

 

Data corrupt or stale; RAM copy of <filename>.

 

The non-volatile RAM copy of a file is either corrupt or is out of date with

 

the EEPROM master copy (if one exists). The system automatically

 

re-initializes the file from EEPROM (if appropriate) or from a default

 

algorithm. A potential cause is a failing backup battery.

-226

Lists not same length

 

Attempted to use LIST structure having individual LISTs of unequal

 

length.

-225

Out of memory

 

The device has insufficient memory to perform the requested operation.

-224

Illegal parameter value

 

Used where exact value, from a list of possibilities, was expected.

Illegal parameter value; Command requires at lease one data value.

A modulation type cannot be created without data.

Illegal parameter value; Total user file bits must be greater than 0.

An attempt was made to select a user file that contains 0 bits. Select another user file that contains at least 1 bit.

Illegal parameter value; Total user file bits must be greater than 0 and smaller than 500000.

The current hardware supports user file size between 1 and 500000 bits. Select a different file with the correct size.

26

Error Messages

Page 28
Image 28
Agilent Technologies E4400-90333 manual During the first power up after downloading new firmware into, 226, Length, 225, 224