-399 to -300: Device-Specific Errors

ESG Family Signal Generators

Out of memory; There is not enough memory to create the modulation. Baseband cannot function.

There is not enough memory to create the modulation map. Reduce the size of any list/sweep, and preset the source to clear up any inconsistencies. If the problem persists report this error to the nearest Agilent Technologies sales or service office.

Out of memory; Object Memory Area

Indicates that memory was exhausted during instrument power-on. Report this error to the nearest Agilent Technologies sales or service office.

Out of memory; Pattern exceeds digital modulation memory size. If a user file is selected, it may be too large.

The user has specified a data file or other generation parameters that would require more data generation pattern memory than is available with the particular data generator installed in the instrument.

Out of memory; PRBS xx/xx

There was not enough memory to apply a scramble to data for the baseband generator. In this case, the scramble is not applied to the generator.

Out of memory; Shape subsystem unable to operate.

There is insufficient working memory available. Report this error to the nearest Agilent Technologies sales or service office.

Out of memory; The table editor cannot function properly until more memory is available.

There is insufficient working memory available. Reduce the size of any sweep lists and try again.

Out of memory; The table editor cannot function properly until more memory is available.

There is insufficient memory for the ARB to generate the requested waveform.

Out of memory; Unable to allocate memory for SCPI response.

There is not enough memory to execute this SCPI query. Reduce the size of any list/sweep, and preset the source to clear up any inconsistencies. If the problem persists report this error to the nearest Agilent Technologies sales or service office.