ESG Family Signal Generators-299 to
| Too much data |
| A legal program data element of block, expression or string type was |
| received that contained more data than the device could handle due to |
| memory or related |
Too much data; The current setting results in too many data points. A new symbol map cannot be generated. CDMA2000 has been turned off.
Due to a hardware limitation, only a certain number of unique data points can be created for the IQ map. CDMA2000 mode turns off when the power combination of all channels exceed that limit. Adjust the channel power level and try again.
Too much data; The current setting results in too many data points. A new symbol map cannot be generated. The IQ Rotation has been restored to previous value.
Due to a hardware limitations, only a certain number of unique data points can be created for the IQ map. IQ rotation is designed to alleviate this problem, but used incorrectly can result in undesirable effects. This error indicates that the current IQ rotation is optimal and cannot be changed.
Too much data; The current setting results in too many data points. A new symbol map cannot be generated. The format has been turned off.
Due to a hardware limitation, only 256 unique data points can be created for the IQ map.
Too much data; The current setting results in too many data points. A new symbol map cannot be generated. The current channel has been restored to its previous type.
Due to a hardware limitations, only a certain number of unique data points can be created for the IQ map. This channel type, together with its power setting, has exceeded that limit. The channel type reverts to its previous type to retain a valid signal. To try a different power setting, turn the channel off, switch to the desired type, change the power setting, and then turn the channel back on.
Too much data; The current setting results in too many data points. A new symbol map cannot be generated. The current channel has been turned off.
Error Messages | 27 |