-299 to -200: Execution ErrorsESG Family Signal Generators

-254

Media full

 

A legal program command or query could not be executed because the

 

media was full. For example, there is was no space left on the disk. The

 

definition of what constitutes full media is device-specific.

 

Media full; Save a state register ignored.

 

The state memory subsystem STATE: did not have enough room to save a

 

register. You must free some memory by deleting a file or register using

 

Catalog. Afterwards, try again.

 

Media full; Save a state register failed. State marked

 

available.

 

The state memory subsystem STATE: did not have enough room to save a

 

register, so the register was lost and is now marked available. You must

 

free some memory by deleting a file or register using Catalog. Afterwards,

 

try again.

 

Media full; Unable to delete saved state from non-volatile

 

memory. No instrument state change.

 

The state memory subsystem STATE: was unable to delete a register. You

 

must free some memory by deleting a file or register using Catalog.

 

Afterwards, try again.

-253

Corrupt media

 

A legal program command or query could not be executed because of

 

corrupt media, for instance a bad disk or incorrect disk format. The

 

definition of what constitutes corrupt media is device-specific.

Corrupt media; <media_name>

A source media (possibly EEPROM) for a data file is corrupt. This error is usually seen in conjunction with errors concerning a certain file.

Corrupt media; Arb file system.

The Arb memory area, used for storing waveform files, is corrupt. The system will clear and reconfigure this memory area automatically. A potential cause is the loss of line power in the middle of a write operation.

Corrupt media; User File System

The main memory area, used for storing instrument states and sequences as well as other data files, is corrupt. The system automatically clears and reconfigures this memory area. Potential causes: failing backup battery; the loss of line power to the instrument during a write operation.

22

Error Messages

Page 24
Image 24
Agilent Technologies E4400-90333 manual 254, Definition of what constitutes full media is device-specific, Try again, 253