201 to 702: Device-Specific Errors

ESG Family Signal Generators

A STATe: file could not be saved due to insufficient space, file corruption, or another related problem.

I/O error; Trailing zero found in <filename>. Fixing...

A compressed state file has a zero at its end. This is a sign of file corruption. The device fixes the problem by concealing the zero such that it no longer triggers an error message. The file may be corrupt or unusable.

I/O error; The user file is too large.

The user file selected for transmission as the DCH data must be less than or equal to 4096 bytes. Select a smaller user file.

I/O error; The user file is too large.

The user file selected for transmission as the TPC pattern must be less than or equal to 256 bytes. Select a smaller user file.

I/O error; Unable to delete saved state from non-volatile memory. No instrument state change.

A STATe: file could not be deleted due to the file not being found, file corruption, or another file-related problem. If the file is displayed by a memory catalog, delete it explicitly.

I/O error; Unable to read bit file.

There is an error reading the specified bit file. Try selecting the file again.

I/O error; Unable to recall from non-volatile memory. No instrument state change.

The state file is not readable and the recall was aborted.

214Not owner; Unable to delete saved state from non-volatile memory. No instrument state change.

The user has attempted to write to a read-only memory subsystem.

218I/O error; filename

An invalid file name was detected during ARB segment or sequence processing.

501Attenuator hold setting over range; Frequency change forced attenuator adjust.

56

Error Messages