ESG Family Signal Generators | 201 to 702: |
DSP code is invalid; Flex DG DSP code is invalid.
The version number of the DSP firmware in the FlexDG data generator is not the version expected by the instrument firmware. The instrument must be upgraded to a matching set of DSP and instrument firmware.
DSP code is invalid; Incorrect demodulator boot code revision.
The demodulator board has an invalid revision of DSP boot code.
DSP code is invalid; Incorrect FPGA revision on Option UN7.
The Option UN7 hardware has an invalid revision of FPGA code.
DSP code is invalid; Running boot code only.
ARB is not functional.
The dual arbitrary waveform generator (Option UND) is running boot code only. The dual arbitrary waveform generator commands will not function. Report this error to the nearest Agilent Technologies sales or service office.
DSP code is invalid; Running boot code only, Demodulator is not functional.
The demodulator DSP ROMs have only boot code loaded: the DSP will not function. Report this error to the nearest Agilent Technologies sales or service office.
DSP code is invalid; Unrecognized version.
Some ARB commands may not function.
The dual arbitrary waveform generator (Option UND) is running an unsupported version of digital signal processor (DSP) code. The dual arbitrary waveform generator may not function correctly. Report this error to the nearest Agilent Technologies sales or service office.
DSP code is invalid; Unrecognized program.
ARB is not functional.
The dual arbitrary waveform generator (Option UND) is running an unknown digital signal processor (DSP) program. The dual arbitrary waveform generator commands will not function. Report this error to the nearest Agilent Technologies sales or service office.
620Clock rate error.
The input clock rate for the bit error rate was too fast. Maximum clock rate is 2 Mbps for the 2 Mbps mode, and 10 Mbps for the 10 Mbps mode.
Error Messages | 67 |