ESG Family Signal Generators

-399 to -300: Device-Specific Errors

 

Out of memory; Unable to check Data Generator memory.

 

There was not enough memory in the system to properly complete the data

 

generator memory test. This does not imply a data generator memory

 

failure. Check all other error messages to identify possible causes,

 

discontinue list/sweep mode to free some memory, and repeat the test.

 

Out of memory; Unable to display format window.

 

The instrument was unable to create part of the graphical user interface

 

due to an inability to allocate memory (possibly due to fragmentation).

 

Report this error to the nearest Agilent Technologies sales or service office.

 

The instrument is still functional.

 

Out of memory; Unable to display protocol window.

 

The instrument was unable to create part of the graphical user interface

 

due to an inability to allocate memory (possibly due to fragmentation).

 

Report this error to the nearest Agilent Technologies sales or service office.

 

The instrument is still functional.

 

Out of memory; Unable to display timeslot window.

 

The instrument was unable to create part of the graphical user interface

 

due to an inability to allocate memory (possibly due to fragmentation).

 

Report this error to the nearest Agilent Technologies sales or service office.

 

The instrument is still functional.

 

Out of memory; Unable to verify instrument state file.

 

An instrument state file could not be accessed and verified because of

 

insufficient memory. Reduce the size of any sweep lists and try again.

 

Out of memory; User burst shape could not be generated.

 

There is insufficient working memory available. Report this error to the

 

nearest Agilent Technologies sales or service office.

-320

Storage fault

 

The firmware detected a fault when using data storage. This error is not

 

an indication of physical damage or failure of any mass storage element.

-315

Configuration memory lost

 

Indicates that non-volatile configuration data saved by the device has been

 

lost. The meaning of this error is device-dependent.

Error Messages