ESG Family Signal Generators

-299 to -200: Execution Errors

Settings conflict; Cannot have more than 4 EDGE multiframe timeslots on simultaneously. The selected timeslot has been turned off.

An unsupported timeslot type was encountered. EDGE does not support the simultaneous activation of more than four multiframe timeslots.

Settings conflict; Cannot have uplink protocols while the control frame is on. Frames 1-17 timeslots changed to downlink.

Indicates that, in TETRA mode, some of the timeslots are set to uplink. When the control frame is turned on in TETRA, all timeslots in all frames must use downlink protocols. The instrument enforces this by reconfiguring the timeslots that are not using downlink protocols when the control frame is switched on.

Settings conflict; Cannot select uplink protocols while the control frame is on. Selection changed to custom.

Indicates that in TETRA mode, an attempt has been made to select uplink protocols while the control frame is switched on. TETRA requires that the control frame be switched off in order to select uplink protocols.

Settings conflict; Carrier formats must be the same.

All digital modulation multicarriers must be the same format.

Settings conflict; Channel #<NUMBER> must be on for any transport channel following it to be on. All transport channels following channel #<NUMBER> have been turned off.

Transport channels must be turned on sequentially. Therefore, all transport channels with a higher number than a channel that is turned off will be automatically turned off.

Settings conflict; Channel #<NUMBER> must be turned on before channel #<NUMBER> can be turned on. Channel #<NUMBER> has been turned off.

Transport channels must be turned on sequentially. For example, channels 1, 2, and 3 must be turned on before channels 4 through 6 can be activated.