At the Configure GPIB menu, press More to go to
an Additional Configure GPIB Menu (below).
Press Pwr Mtr Address to change the address of the
power meter on the GPIB (the power meter’s default
address is 13). Enter the new address, between 1
and 30, using the cursor control key or the data en-
try key pad and the terminator key
Hz
ns
ADRS
The new GPIB address will appear on the display.
Press Pwr Mtr Select to select the power meter
model being used. (Supported power meters are the
Anritsu ML2437A, ML2438A, and ML4803Aand
Hewlett-Packard 437B, 438A, and 70100A.)
Press Previous Menu to return to the main Config-
ure GPIB Menu display.
At the Configure GPIB menu, press Previous Menu
to return to the System Configuration Menu display.
Creating a Power-Offset Table
The 681XXC must be in CW frequency mode and
fixed (non-swept) power level mode in order to cre-
ate a power-offset table for user level flatness cor-
rection.
Place the signal generator in CW frequency mode by
pressing the main menu key
CW/SWEEP
SELECT
At the resulting menu display,press CW . The
681XXC is now in CW frequency mode.
681XXC OM 3-69LOCAL (FRONT LEVELING
PANEL) OPERATION OPERATIONS