If the address of the 681XXC on the System Bus
needs changing, press GPIB Address . Enter the
new address using the cursor control key or the data
entry keypad and the terminator key
HZ
ms
ADRS
The new GPIB address will appear on the display.
Tochange the data terminator, press Line Term to
select the correct GPIB data delimiter.
Toturn SS mode off, press SS Mode .
The signal generator is now configured for 360B
tracking mode operation.
Initiating 360B TrackingMode Operations
Turnon the 360B and configure it for tracking mode
of operation. (Refer to the 360B VNAoperation man-
ual.) Once configured, the 360B should take control
of the signal generator.
When the 360B takes control, the display of all pa-
rameters on the 681XXC is disabled and the mes-
sages Secure Mode Active and Remote appear on the
front panel display.
Terminating360B Tracking Mode Operations
Toterminate 360B VNA tracking mode operations,
you must first return the 681XXC to local control
and then turn off the Secure mode.
Toreturn the 681XXC to local control, turn off the
360B VNA.
On the 681XXC, press SYSTEM , then Reset . This
turns off the Secure mode.
681XXC OM 7-15
USE WITH USE WITH A 360B VECTOR
OTHER INSTRUMENTS NETWORK ANALYZER