Press VNAMenu to go to the VNAMenu display
(below).
This menu lets you set the frequency offset and out-
put power level for the Slave instrument in the VNA
mode.
Press VNAOffset to open the slave frequency offset
parameter.
Press VNALevel to open the slave output power
level parameter.
Open the parameter you wish to change, then edit
the current value using the cursor control key or ro-
tary data knob or enter a new value using the key
pad and appropriate termination key.When you
have finished setting the open parameter,close it by
pressing its menu soft-key or by making another
menu selection.
Press Previous Menu toreturn to the Slave Fre-
quencies List Menu display.
Return to the Master-Slave menu and press Enable
to begin master-slave operation.
7-8 681XXC OMUSE WITH MASTER-SLAVE
OTHER INSTRUMENTS OPERATION
SLAVE
During master-slave operations in
VNA mode, this error message is
displayed on the Master
instrument when the slave offset
value entered results in a CW fre-
quency or frequency sweep outside
the range of the slave unit. Enter-
ing a valid offset value clears the
error.