Step 4 Connect the Master instrument RF OUT-
PUT and the Slave instrument RF OUT-
PUT to the appropriate connections on the
DUT.
Initiating
Master-Slave
Operation
The following paragraphs describe how to set up
both instruments to perform master-slave opera-
tions. Use the CW Frequency Mode menu map
(Chapter 4, Figure 4-2) to follow the menu se-
quences.
Toinitiate master-slave operation, turn on both
instruments and place them in CW mode. The CW
Menu (below) is displayed.
On the Master instrument, press Master Slave to
go to the Master-Slave Menu display (below).
This menu lets you perform the following:
qGo to the Slave Frequencies List menu.
qSet the dF frequency for the Slave unit.
qSet the Slave unit’smain power level (L1).
qSet the alternate sweep power level (L2) for
the Slave unit.
qTurnmaster-slave operation on and off.
Press Slave Freqs to go to the Slave Frequencies
List Menu display (next page).
681XXC OM 7-5
USE WITH MASTER-SLAVE
OTHER INSTRUMENTS OPERATION
NOTE
Master-slaveoperations are always
initiated in the CW frequency
mode. Once initiated, you then can
change to a sweep frequency mode
of operation by selecting the de-
siredfrequency mode on the Master
instrument.