This menu lets you edit the listed frequencies for
the Slave instrument.
Use the cursor control key to select a frequency pa-
rameter from the list, then press Edit to edit its
value. Edit the current frequency parameter value
using the cursor control key or rotary data knob or
enter a new value using the key pad and approriate
termination key.Press Edit again to close the open
frequency parameter.
When you are finished editing the slave frequencies,
press Previous Menu to return to the Master-Slave
Menu (below).
The Master-Slave menu lets you set the dF fre-
quency and L1 and L2 power level parameters for
the Slave unit.
Press Slave DF to open the dF frequency parame-
ter.
Press Slave L1 to open the main power level pa-
rameter.
Press Slave L2 to open the alternate sweep power
level parameter.
Open the parameter you wish to change, then edit
the current value using the cursor control key or ro-
tary data knob or enter a new value using the key
7-6 681XXC OM
USE WITH MASTER-SLAVE
OTHER INSTRUMENTS OPERATION
NOTE
Upon reset, the slave frequencies
(F0- F9 and M0 - M9) return to the
default values shown here.