7-2 MASTER-SLAVE
OPERATION Master-slaveoperation consists of connecting any two 68XXXC and/or
68XXXB instruments together and configuring them so that they pro-
duce CW and synchronized, swept output signals at an operator-
selectable frequency offset. One instrument (the Master) controls the
other (the Slave) via interface cables between their rear panel AUX I/O
and SERIALI/O connectors. The two units are phase-locked together
by connecting them to the same 10 MHz reference time base.
Connecting
the Instru-
ments
Connect the two instruments, shown in Figure 7-1,
as follows:
Step 1 Connect the 3-port AUX I/O cable end la-
beled “MASTER” to the rear panel AUX
I/O connector on the Master instrument.
Connect the AUX I/O cable labeled
“SLAVE”to the rear panel AUX I/O con-
nector on the Slave instrument.
Step 2 Connect the ends of the flat interface ca-
ble to the rear panel Serial I/O connectors
on the Master and Slave instruments.
Step 3 Connect one end of a coaxial cable to the
rear panel 10 MHz REF OUT connector on
the Master instrument. Connect the other
end to the rear panel 10 MHz REF IN con-
nector on the Slave instrument.

7-4 681XXC OM

USE WITH MASTER-SLAVE
OTHER INSTRUMENTS OPERATION
Figure 7-1. 68XXXB/68XXXC Configuration for Master-Slave Operation
NOTES
When connecting two instruments
together for Master-Slave opera-
tions, always use a Anritsu
Master-Slave interface cable set,
Part No. ND36329.
If a Model 56100A Scalar Network
Analyzer is being used with the
master-slaveconfiguration, (1) con-
nect the AUX I/O cable end labeled
“SNA” to the rear panel AUX I/O
connector on the 56100A SNAand
(2) connect a dedicated system bus
cable (P/N 2100-1) between the
Master instrument rear panel
IEEE-488 GPIB connector and the
56100A SNA rear panel DEDI-
CATEDGPIB connector.