2.6Grounding the flowmeter components

Installing the Transmitter

2.6Grounding the flowmeter components

Grounding requirements depend on the installation type (see Figure 2-1).Grounding methods for each flowmeter component are listed in Table 2-3.

InstallationarchitectureComponents CAUTION

Improper grounding could cause measurement error.

To reduce the risk of measurement error:

Ground the transmitter to earth, or follow ground network requirements for the facility.

For installation in an area that requires intrinsic safety, refer to Micro Motion approval documentation, shipped with the transmitter or available from the Micro Motion web site.

For hazardous area installations in Europe, refer to standard EN 60079-14 if national standards do not apply.

If national standards are not in effect, follow these grounding guidelines:

Use copper wire, 14 AWG (2,5 mm2) or larger wire size, for grounding.

Keep all ground leads as short as possible, less than 1 Ω impedance.

Connect ground leads directly to earth, or follow plant standards.

Table 2-3Grounding methods for flowmeter components

Installation

 

 

architecture

Components

Grounding method

 

 

 

Integral

Sensor / core processor /

Ground via piping, if possible (see sensor documentation). Otherwise,

 

transmitter

ground according to applicable local standards using either the

 

 

transmitter’s internal or external ground screw.

 

 

 

4-wire remote

Sensor / core processor

See sensor documentation.

 

assembly

 

 

 

 

 

Transmitter

Ground according to applicable local standards, using either the

 

 

transmitter’s internal or external ground screw.

 

 

 

9-wire remote

Sensor / junction box

See sensor documentation.

 

 

 

 

Transmitter / core

Ground according to applicable local standards, using either the

 

processor assembly

transmitter’s internal or external ground screw, or the core processor’s

 

 

internal ground screw.

 

 

 

Remote core

Sensor

See sensor documentation.

processor with

 

 

Core processor

Ground according to applicable local standards, using either the internal

remote

 

or external ground screw.

transmitter

 

 

 

 

Transmitter

Ground according to applicable local standards, using either the

 

 

transmitter’s internal or external ground screw.

 

 

 