Emerson 2700, 1700 Rmax = Vsupply - 12/0.023, Output Wiring - Intrinsically Safe, Specifications

Models: 2700 1700

1 80
Download 80 pages 1.45 Kb
Page 39
Image 39
Rmax = (Vsupply – 12)/0.023

Output Wiring – Model 1700/2700 Intrinsically Safe Transmitters

Figure 5-2Safe area mA output load resistance values

Rmax = (Vsupply – 12)/0.023

If communicating with HART, a minimum of 250 Ω and 17.5 V is required

 

1000

 

 

 

 

 

 

 

 

 

 

900

 

 

 

 

 

 

 

 

 

 

800

 

 

 

 

 

 

 

 

 

(Ohms)

700

 

 

 

 

 

 

 

 

 

600

 

 

 

 

 

 

 

 

 

load

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

R

500

 

 

 

 

 

 

 

 

 

resistor

 

 

 

 

 

 

 

 

 

400

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

External

300

 

 

 

 

OPERATING REGION

 

 

 

 

 

 

 

 

 

200

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

100

 

 

 

 

 

 

 

 

 

 

0

 

 

 

 

 

 

 

 

 

 

12

14

16

18

20

22

24

26

28

30

Supply voltage VDC (Volts)

Figure 5-3Safe area HART/analog single-loop wiring

 

 

 

+

 

VDC

+

 

 

 

 

 

 

mA1

 

 

 

HART-

 

 

 

compatible host

Rload

 

or controller

 

 

(250–600

Ω

 

 

resistance)

 

 

Note: See Figure 5-2for voltage and resistance values.

Output Wiring – Intrinsically Safe

Output Wiring – Configurable I/O Output Wiring – Fieldbus/PROFIBUS

Specifications

Installation Manual

35

Page 39
Image 39