
Output Wiring – Model 1700/2700 Analog Transmitters
Note: If you will configure the transmitter to poll an external temperature or pressure device, you must wire the mA output to support HART communications. You may use either HART/analog
It is the user’s responsibility to verify that the specific installation meets the local and national safety requirements and electrical codes.
Figure 4-1 Basic analog wiring
+
–
mA output loop
820 Ω maximum loop resistance
+ |
– |
00042 |
Frequency receiving device
Output voltage level is +24 VDC ± 3%
30 | Micro Motion® Model 1700 and 2700 Transmitters |