Wiring the Transmitter to the Sensor

Figure 3-7Wrapping the shield drain wires

Before You Begin

e.Place the shielded heat shrink over the exposed shield drain wire(s). The tubing should completely cover the drain wires. See Figure 3-8.

f.Without burning the cable, apply heat (250 °F or 120 °C) to shrink the tubing.

Figure 3-8Applying the heat shrink

g.Position gland clamping insert so the interior end is flush with the heat shrink.

h.Fold the cloth shield or braid and drain wires over the clamping insert and approximately 1/8 inch (3 mm) past the O-ring. See Figure 3-9.

Figure 3-9Folding the cloth shield

i.Install the gland body into the core processor housing conduit opening. See Figure 3-10.

Installing the Transmitter

Sensor Wiring

Output Wiring – Analog

Installation Manual

25

Page 29
Image 29
Emerson 2700, 1700 installation manual 7Wrapping the shield drain wires