Emerson manual Review Configuration Data, CHECK OUTPUT Process Variables, Rosemount 3051N

Models: 3051N

1 100
Download 100 pages 43.51 Kb
Page 19
Image 19
REVIEW CONFIGURATION DATA

Reference Manual

00809-0100-4808, Rev CA June 2008

Rosemount 3051N

REVIEW CONFIGURATION DATA

HART Comm

1, 5

 

 

CHECK OUTPUT

Process Variables

HART Comm.

2

 

 

NOTE

Information and procedures in this section that make use of HART Communicator fast key sequences assume that the transmitter and communicator are connected, powered, and operating correctly. If you are not familiar with the HART Communicator or fast-key sequences, refer to Appendix A: HART Communicator.

Before you place the transmitter into operation, it is recommended that you review the transmitter configuration data that was set at the factory. You should review the following configuration data:

Transmitter Model

Type

Tag

Range

Date

Descriptor

Message

Minimum and Maximum

 

Sensor Limits

Minimum Span

Units

4 and 20 mA points

Output (linear or sq. root)

Damping

Alarm Setting (high, low)

Security Setting (on, off)

Local Zero/Span Keys

 

(enabled, disabled)

Integral Meter

Sensor Fill

Isolator Material

Flange (type, material)

O-Ring Material

Drain/Vent

Remote Seal (type, fill fluid,

Transmitter S/N

isolator material, number)

 

Address

Sensor S/N

Before performing other transmitter on-line operations, review the digital output parameters to ensure that the transmitter is operating properly and is configured to the appropriate process variables.

The process variables for the Rosemount 3051N provide the transmitter output, and are continuously updated. The process variable menu displays the following process variables:

Pressure

Percent of Range

Analog Output

The pressure reading in both Engineering Units and Percent of Range will continue to track with pressures outside of the defined range from the lower to the upper range limit of the sensor module.

2-7

Page 19
Image 19
Emerson manual Review Configuration Data, CHECK OUTPUT Process Variables, Rosemount 3051N, Reference Manual