Emerson manual Fast Key Sequences Hot Key options DIAGNOSTIC, Messages, Rosemount 3051N

Models: 3051N

1 100
Download 100 pages 43.51 Kb
Page 90
Image 90
Fast Key Sequences

Rosemount 3051N

Reference Manual

00809-0100-4808, Rev CA June 2008

Fast Key Sequences

Hot Key options

DIAGNOSTIC

MESSAGES

The Fast Key sequence is a sequence of numerical button presses, corresponding to the menu options that lead you to a given task. Use Table A-1, an alphabetical listing of most on-line functions, to find the corresponding HART fast key sequences. These codes are applicable only to Rosemount 3051N transmitters and the HART Communicator.

The Hot Key menu is a user-definable menu that can store shortcuts for up to 20 of your most frequently performed tasks. For example, if you change device tags, and damping often, you can add options for these functions to the Hot Key menu. Once you are Online, the Hot Key will automatically appear in the toolbar. To add a custom option to the Hot Key menu, refer to the Rosemount 375 Communicator user manual.

To use a Hot Key option, you must first properly connect the Rosemount 375 Field Communicator to a device. You can access the Hot Key menu from any online window. To use a Hot Key option:

1.Connect the Rosemount 375 Field Communicator to a HART loop or device.

2.Tap the Hot Key. The Hot Key menu is displayed.

3.Double-tap the desired option.

The following pages contain a list of messages used by both of the Rosemount 275 and Rosemount 375 HART Communicators (HC) and their corresponding descriptions.

Variable parameters within the text of a message are indicated with <variable parameter>.

Reference to the name of another message is identified by [another message].

Message

 

Description

 

 

 

1k snsr EEPROM

 

Return to Rosemount Nuclear Instruments, Inc. for repair.

error-factory ON

 

 

 

 

 

1k snsr EEPROM

 

Use the HART communicator to reset the following parameters:

error-user-no out ON

 

remote seal isolator, remote seal fill fluid, flange material, o-ring

 

 

material, transmitter type, remote seal type, flange type, meter

 

 

type, number of remote seals.

 

 

 

1k snsr EEPROM

 

Perform a full trim to recalibrate the transmitter.

error-user ON

 

 

 

 

 

4k micro EEPROM

 

Return to Rosemount Nuclear Instruments, Inc. for repair.

error-factory ON

 

 

 

 

 

4k micro EEPROM

 

Use the HART communicator to reset the message field.

error-user-no out ON

 

 

 

 

 

4k micro EEPROM

 

Use the HART communicator to reset the following parameters:

error-user ON

 

units, range values, damping, analog output, transfer function,

 

 

tag, scaled meter values. Perform a d/a trim to ensure that the

 

 

error is corrected.

 

 

 

4k snsr EEPROM

 

Return to Rosemount Nuclear Instruments, Inc. for repair.

error-factory ON

 

 

 

 

 

4k snsr EEPROM

 

Use the HART communicator to reset the temperature units

error-user ON

 

and the calibration type.

 

 

 

A-14